Pediatric chest CT at 70 kVp: a feasibility study in 129 children.
Niemann, Tilo; Henry, Simon; Duhamel, Alain; Faivre, Jean-Baptiste; Deschildre, Antoine; Colas, Lucie; Santangelo, Teresa; Remy, Jacques; Remy-Jardin, Martine.
Afiliação
  • Niemann T; Department of Thoracic Imaging, Hospital Calmette, Univ Lille Nord de France, Lille, France.
Pediatr Radiol ; 44(11): 1347-57, 2014 Nov.
Article em En | MEDLINE | ID: mdl-24925355
ABSTRACT

BACKGROUND:

Before introducing 70-kVp settings in the low-kilovoltage strategies for pediatric examinations, it was mandatory to demonstrate, at similar dose levels, an equivalence of image quality at 70 kVp and 80 kVp.

OBJECTIVE:

To assess image quality of chest CT examinations acquired at 70 kVp in comparison with standard scanning at 80 kVp. MATERIALS AND

METHODS:

We prospectively evaluated 129 children with a 70-kVp scanning protocol (group 1). All scanning parameters were kept similar to those usually selected for pediatric standard 80-kVp protocols, except the milliamperage increased by a factor of 1.6 to maintain comparable radiation dose. Image quality of group 1 examinations was compared to that of a paired population scanned at 80 kVp (group 2). The noninferiority hypothesis was fixed at 10% of the mean level of image noise.

RESULTS:

There was no significant difference in the mean dose length product (DLP) and the volume computed tomography dose index (CTDIvol) between the groups (DLP 20.5 ± 5.8 mGy.cm [group 1] vs. 19.7 ± 7.6 mGy.cm [group 2]; P = 0.06) (CTDIvol 0.8 ± 0.1 mGy [group 1] vs. 0.8 ± 0.18 mGy [group 2]; P = 0.94). The mean of differences in image noise between group 1 and group 2 examinations was -1.38 (-2.59; -0.18), verifying the noninferiority hypothesis. Subjective image quality did not significantly differ between group 1 and group 2 examinations (P = 0.18).

CONCLUSION:

At equivalent radiation dose levels, 70-kVp protocols provide similar image quality to that achievable at 80 kVp.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doses de Radiação / Proteção Radiológica / Radiografia Torácica / Tomografia Computadorizada por Raios X Idioma: En Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doses de Radiação / Proteção Radiológica / Radiografia Torácica / Tomografia Computadorizada por Raios X Idioma: En Ano de publicação: 2014 Tipo de documento: Article