Your browser doesn't support javascript.
loading
Two case reports of retained steel insulin pump infusion set needles.
Plager, Phillip; Murati, Michael A; Moran, Antoinette; Sunni, Muna.
Afiliação
  • Plager P; Department of Pediatrics, University of Minnesota Masonic Children's Hospital, Minneapolis, MN, 55454, USA.
  • Murati MA; Department of Pediatrics, University of Minnesota Masonic Children's Hospital, Minneapolis, MN, 55454, USA.
  • Moran A; Department of Pediatrics, University of Minnesota Masonic Children's Hospital, Minneapolis, MN, 55454, USA.
  • Sunni M; Department of Pediatrics, University of Minnesota Masonic Children's Hospital, Minneapolis, MN, 55454, USA.
Pediatr Diabetes ; 17(2): 160-3, 2016 Mar.
Article em En | MEDLINE | ID: mdl-25683199
ABSTRACT
Insulin pumps are common in the management of type 1 diabetes (T1D). We report two cases of metal insulin infusion set needles which broke off the tubing and remained embedded in the soft tissue of two boys with T1D (five needles in one case, and one needle in the other). The patient with five retained needles was asymptomatic and had a normal physical examination, and the missing needles were only detected using pelvic X-ray; the second patient had only mild discomfort. While these are the first such cases reported in the medical literature, there may be other cases which have gone unnoticed, suggesting the potential need to explore the safety of this product further.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Sistemas de Infusão de Insulina / Diabetes Mellitus Tipo 1 / Corpos Estranhos / Insulina / Agulhas Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Sistemas de Infusão de Insulina / Diabetes Mellitus Tipo 1 / Corpos Estranhos / Insulina / Agulhas Idioma: En Ano de publicação: 2016 Tipo de documento: Article