Preliminary evaluation of an improved enzymatic assay method for measuring potassium concentrations in serum.
Ota, Eri; Sakasegawa, Shin-ichi; Ueda, Shigeru; Konishi, Kenji; Akimoto, Masaru; Tateishi, Takiko; Kawano, Miki; Hokazono, Eisaku; Kayamori, Yuzo.
Afiliação
  • Ota E; Division of Biological Science and Technology, Department of Health Science, Faculty of Medical Sciences, Kyushu University, Japan.
  • Sakasegawa S; Asahi Kasei Pharma Corporation, Japan. Electronic address: sakasegawa.sb@om.asahi-kasei.co.jp.
  • Ueda S; Asahi Kasei Pharma Corporation, Japan.
  • Konishi K; Asahi Kasei Pharma Corporation, Japan.
  • Akimoto M; Division of Biological Science and Technology, Department of Health Science, Faculty of Medical Sciences, Kyushu University, Japan.
  • Tateishi T; Division of Biological Science and Technology, Department of Health Science, Faculty of Medical Sciences, Kyushu University, Japan.
  • Kawano M; Division of Biological Science and Technology, Department of Health Science, Faculty of Medical Sciences, Kyushu University, Japan.
  • Hokazono E; Division of Biological Science and Technology, Department of Health Science, Faculty of Medical Sciences, Kyushu University, Japan.
  • Kayamori Y; Division of Biological Science and Technology, Department of Health Science, Faculty of Medical Sciences, Kyushu University, Japan.
Clin Chim Acta ; 446: 73-5, 2015 Jun 15.
Article em En | MEDLINE | ID: mdl-25892675
BACKGROUND: K(+) has important physiological functions. K(+) concentrations in serum are generally determined using ion-selective electrodes (ISEs), though measurement using reagents in aqueous medium is also useful. METHODS: K(+) concentrations were measured using recombinant inosine 5'-monophosphate dehydrogenase (IMPDH), which was activated only by K(+) and NH4(+). Exogenous NH4(+) and endogenous NH4(+) were eliminated using glutamine synthase. RESULTS: Regression analysis of the enzymatic assay (y) vs. the ISE method (x) gave the following relation: y=1.03x+0.09 (n=54, Sy,x=0.06 mmol/l). The linear range was up to 12 mmol/l when 1 U/ml IMPDH was used. CONCLUSION: Advantages of the proposed assay method are: (i) the measured range is wider than that of existing enzymatic methods; (ii) the conditions for K(+) determination can be maintained constant, regardless of the amount of NH4(+) in the analyte and reagents; and (iii) the elimination system is simpler because the recombinant IMPDH is stimulated by only K(+) and NH4(+) and is unaffected by biological materials.