What effect do different 200 µm laser fibres have on deflection and irrigation flow rates in a flexible ureterorenoscope?
Wright, Anna E; Williams, Kevin; Rukin, Nicholas J.
Afiliação
  • Wright AE; Department of Urological Surgery, The Royal Wolverhampton Hospital, Wolverhampton Road, Wolverhampton, WV10 0QP, UK, annawright88@doctors.org.uk.
Lasers Med Sci ; 30(5): 1565-8, 2015 Jul.
Article em En | MEDLINE | ID: mdl-25967119
ABSTRACT
The objective of the study is to evaluate the reduction in flow and scope deflection of four leading 200-µm marketed laser fibres (Boston Scientific Flexiva™ 200, Boston Scientific Flexiva™ Trac Tip 200, Lumenis SlimLine™ EZ200 and Optical Integrity ScopeSafe™) via a flexible ureterorenoscope. A laboratory-based bench test was performed using a Flex X2™ flexible ureterorenoscope. Mean upward/downward deflection angles and flow rates (ml/min) for each fibre were calculated and compared to a control. The Optical Integrity ScopeSafe™ fibre has the least loss of deflection, losing only 8 % upward and 6 % downward deflection. Deflection loss was significantly less with this fibre compared to all other fibres (p < 0.0001). Mean flow rates were significantly greater with the Optical Integrity ScopeSafe™ laser fibre at 23 ml/min (p < 0.0001). Despite all fibres marketed as 200 µm, the deflection and flow properties show marked variations. The Optical Integrity ScopeSafe™ 200-µm laser fibre offers the best overall performance with significantly improved flow rates and the least loss of scope deflection.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Ureteroscópios / Fibras Ópticas Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Ureteroscópios / Fibras Ópticas Idioma: En Ano de publicação: 2015 Tipo de documento: Article