Your browser doesn't support javascript.
loading
Validity and Predictive Value of a Portable Two-Channel Sleep-Screening Tool in the Identification of Sleep Apnea in Patients With Heart Failure.
de Vries, Grietje E; van der Wal, Haye H; Kerstjens, Huib A M; van Deursen, Vincent M; Stegenga, Boudewijn; van Veldhuisen, Dirk J; van der Hoeven, Johannes H; van der Meer, Peter; Wijkstra, Peter J.
Afiliação
  • de Vries GE; Department of Pulmonary Diseases, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands; GRIAC Research Institute, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • van der Wal HH; Department of Cardiology, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • Kerstjens HA; Department of Pulmonary Diseases, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands; GRIAC Research Institute, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • van Deursen VM; Department of Cardiology, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • Stegenga B; Department of Oral and Maxillofacial Surgery, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • van Veldhuisen DJ; Department of Cardiology, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • van der Hoeven JH; Department of Clinical Neurophysiology, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
  • van der Meer P; Department of Cardiology, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands. Electronic address: p.van.der.meer@umcg.nl.
  • Wijkstra PJ; Department of Pulmonary Diseases, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands; GRIAC Research Institute, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands; Center for Home Mechanical Ventilation, University of Gronin
J Card Fail ; 21(10): 848-55, 2015 Oct.
Article em En | MEDLINE | ID: mdl-26095313
ABSTRACT

BACKGROUND:

Sleep apnea is an important comorbidity in heart failure (HF) and is associated with an adverse outcome. Diagnosing sleep apnea is difficult, and polysomnography, considered to be the criterion standard, is not widely available. We assessed the validity of a portable 2-channel sleep-screening tool for the identification of sleep apnea in patients with HF. METHODS AND

RESULTS:

One hundred patients with stable HF had simultaneous recordings of home-based polysomnography and the screening tool (Apnealink). To compare the apnea-hypopnea index of the screening tool with polysomnography, intraclass correlation (ICC), sensitivity, and specificity were calculated, and a Bland-Altman plot and receiver operating characteristic (ROC) curves were constructed. Ninety valid measurements with the screening tool were obtained (mean age 65.5 ± 11.0 y, 72% male, mean left ventricular ejection fraction 34.6 ± 11.0%). Agreement between the screening tool and polysomnography was high (ICC 0.85). The optimal cutoff value was apnea-hypopnea index ≥15/h (area under the ROC curve 0.94). Sensitivity and specificity were 92.9% and 91.9%, respectively.

CONCLUSIONS:

The screening tool is useful in excluding the presence of sleep apnea in HF patients to refer only high-risk patients for more extensive polysomnography. This method may potentially reduce the need for the more expensive polysomnography.