An iterative deconvolution algorithm for image recovery in clinical CT: A phantom study.
Slavine, Nikolai V; Guild, Jeffrey; McColl, Roderick W; Anderson, Jon A; Oz, Orhan K; Lenkinski, Robert E.
Afiliação
  • Slavine NV; Translational Research, Department of Radiology, UT Southwestern Medical Center, 5323 Harry Hines Boulevard, Dallas, TX, 75390-9061, USA. Electronic address: Nikolai.Slavine@UTSouthwestern.edu.
  • Guild J; Clinical Medical Physics, Department of Radiology, UT Southwestern Medical Center, 5323 Harry Hines Boulevard, Dallas, TX 75390-9061, USA.
  • McColl RW; Clinical Medical Physics, Department of Radiology, UT Southwestern Medical Center, 5323 Harry Hines Boulevard, Dallas, TX 75390-9061, USA.
  • Anderson JA; Clinical Medical Physics, Department of Radiology, UT Southwestern Medical Center, 5323 Harry Hines Boulevard, Dallas, TX 75390-9061, USA.
  • Oz OK; Nuclear Medicine, Department of Radiology, UT Southwestern Medical Center, 5323 Harry Hines Boulevard, Dallas, TX 75390-9061, USA.
  • Lenkinski RE; Translational Research, Department of Radiology, UT Southwestern Medical Center, 5323 Harry Hines Boulevard, Dallas, TX, 75390-9061, USA.
Phys Med ; 31(8): 903-911, 2015 Dec.
Article em En | MEDLINE | ID: mdl-26143585
ABSTRACT

PURPOSE:

To study the feasibility of using an iterative reconstruction algorithm to improve previously reconstructed CT images which are judged to be non-diagnostic on clinical review. A novel rapidly converging, iterative algorithm (RSEMD) to reduce noise as compared with standard filtered back-projection algorithm has been developed. MATERIALS AND

METHODS:

The RSEMD method was tested on in-silico, Catphan(®)500, and anthropomorphic 4D XCAT phantoms. The method was applied to noisy CT images previously reconstructed with FBP to determine improvements in SNR and CNR. To test the potential improvement in clinically relevant CT images, 4D XCAT phantom images were used to simulate a small, low contrast lesion placed in the liver.

RESULTS:

In all of the phantom studies the images proved to have higher resolution and lower noise as compared with images reconstructed by conventional FBP. In general, the values of SNR and CNR reached a plateau at around 20 iterations with an improvement factor of about 1.5 for in noisy CT images. Improvements in lesion conspicuity after the application of RSEMD have also been demonstrated. The results obtained with the RSEMD method are in agreement with other iterative algorithms employed either in image space or with hybrid reconstruction algorithms.

CONCLUSIONS:

In this proof of concept work, a rapidly converging, iterative deconvolution algorithm with a novel resolution subsets-based approach that operates on DICOM CT images has been demonstrated. The RSEMD method can be applied to sub-optimal routine-dose clinical CT images to improve image quality to potentially diagnostically acceptable levels.