Your browser doesn't support javascript.
loading
Simple method for reducing the first-order optical nonlinearity in a heterodyne laser interferometer.
Appl Opt ; 54(20): 6321-6, 2015 Jul 10.
Article em En | MEDLINE | ID: mdl-26193410
ABSTRACT
A simple method was proposed by using a tunable attenuator fitted in the reference or measurement arm of a heterodyne laser interferometer to adjust the values of mixing laser beams while the spectrum of the measurement signal is monitored using a signal analyzer. The effectiveness of the proposed method in reducing the first-order optical nonlinearity was verified through experiments. Results indicated that the peak value of the first-order optical nonlinearity could be reduced from 5.15 to 0.24 nm. It was therefore concluded that the proposed method was applicable to ultraprecision laser interferometry.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2015 Tipo de documento: Article