Your browser doesn't support javascript.
loading
Hybrid deterministic-stochastic modeling of x-ray beam bowtie filter scatter on a CT system.
Liu, Xin; Hsieh, Jiang.
Afiliação
  • Liu X; Missouri University of Science and Technology, Nuclear Engineering, Rolla, MO, USA.
  • Hsieh J; GE Healthcare, Waukesha, WI, USA.
J Xray Sci Technol ; 23(5): 593-600, 2015.
Article em En | MEDLINE | ID: mdl-26409426
Knowledge of scatter generated by bowtie filter (i.e. x-ray beam compensator) is crucial for providing artifact free images on the CT scanners. Our approach is to use a hybrid deterministic-stochastic simulation to estimate the scatter level generated by a bowtie filter made of a material with low atomic number. First, major components of CT systems, such as source, flat filter, bowtie filter, body phantom, are built into a 3D model. The scattered photon fluence and the primary transmitted photon fluence are simulated by MCNP - a Monte Carlo simulation toolkit. The rejection of scattered photon by the post patient collimator (anti-scatter grid) is simulated with an analytical formula. The biased sinogram is created by superimposing scatter signal generated by the simulation onto the primary x-ray beam signal. Finally, images with artifacts are reconstructed with the biased signal. The effect of anti-scatter grid height on scatter rejection are also discussed and demonstrated.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia Computadorizada por Raios X Idioma: En Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Tomografia Computadorizada por Raios X Idioma: En Ano de publicação: 2015 Tipo de documento: Article