Chemiluminometric Immunosensor for High-Sensitivity Cardiac Troponin I Employing a Polymerized Enzyme Conjugate as a Tracer.
Lim, Guei-Sam; Seo, Sung-Min; Paek, Sung-Ho; Kim, Seung-Wan; Jeon, Jin-Woo; Kim, Dong-Hyung; Cho, Il-Hoon; Paek, Se-Hwan.
Afiliação
  • Lim GS; Department of Bio-Microsystem Technology, Korea University, Sungbuk-gu, Seoul 136-701, Korea.
  • Seo SM; Devices and Materials Laboratory, LG Electronics Advanced Research Institute, Seocho-gu, Seoul 137-724, Korea.
  • Paek SH; Department of Bio-Microsystem Technology, Korea University, Sungbuk-gu, Seoul 136-701, Korea.
  • Kim SW; Department of Bio-Microsystem Technology, Korea University, Sungbuk-gu, Seoul 136-701, Korea.
  • Jeon JW; Department of Bio-Microsystem Technology, Korea University, Sungbuk-gu, Seoul 136-701, Korea.
  • Kim DH; Department of Bio-Microsystem Technology, Korea University, Sungbuk-gu, Seoul 136-701, Korea.
  • Cho IH; Department of Bio-Microsystem Technology, Korea University, Sungbuk-gu, Seoul 136-701, Korea.
  • Paek SH; Department of Biomedical Laboratory Science, Eulji University, Seongnam, Gyeonggi-do 461-713, Korea.
Sci Rep ; 5: 14848, 2015 Oct 07.
Article em En | MEDLINE | ID: mdl-26442606
ABSTRACT
To detect high-sensitivity cardiac troponin I (hs-cTnI; <0.01 ng/mL) at points of care, we developed a rapid immunosensor by using horseradish peroxidase polymerized in 20 molecules on average (Poly-HRP) as a tracer conjugated with streptavidin (SA-Poly-HRP). As shown in the conventional system, enhanced sensitivity could be achieved by using a sequential binding scheme for the complex formation to contain the huge molecular tracer. We used a 2-dimensional chromatographic technology to carry out the sequential bindings in cross-flow directions. After the complex formation of antigen-antibody with analyte in a vertical direction, SA-Poly-HRP was horizontally supplied across the membrane strip for additional binding via a biotin-SA linkage. The HRP substrate was subsequently supplied along the same direction to produce a chemiluminometric signal, which was measured by a cooled charge-coupled device. Hs-cTnI analysis was completed in this format within 25 min, and the results showed a high correlation with those of the CentaurXP® reference system (R(2) > 0.99). The detection limit of the rapid immunosensor was 0.003 ± 0.001 ng/mL cTnI, corresponding to a 10-fold improvement compared to results using the plain enzyme tracer. This demonstrated the measurement of hs-cTnI in a much more cost-effective manner compared to the automated versions currently available.