Performance demonstration of 4πß(LS)-γ coincidence counting system for standardization of radionuclides with complex decay scheme.
Appl Radiat Isot
; 108: 24-29, 2016 Feb.
Article
em En
| MEDLINE
| ID: mdl-26678524
ABSTRACT
A standardization of (134)Cs and (131)I was carried out in order to demonstrate the performance and applicability of the 4πß(LS)-γ coincidence counting system for standardization of radionuclides with complex decay scheme. The coincidence analyzer, capable of analyzing coincidence between beta and two gamma windows simultaneously, was developed and used for the standardization. The use of this dual coincidence analyzer has reduced the total experimental time by half. The activity concentrations obtained using the 4πß(LS)-γ coincidence counting system, a 4πß(PC)-γ coincidence counting system, and the CIEMAT/NIST method are in excellent agreement with each other within uncertainty limits and hence demonstrates its performance for standardization of radionuclides decaying with complex decay scheme. Hence use of this 4πß(LS)-γ coincidence counting system can be an alternative method suitable to standardize radionuclides with complex decay scheme with acceptable precision.