An analytical comparison between point-of-care uric acid testing meters.
Expert Rev Mol Diagn
; 16(3): 373-82, 2016.
Article
em En
| MEDLINE
| ID: mdl-26689648
OBJECTIVE: A key goal in gout treatment is achieving sustained serum uric acid (sUA) lowering. Point-of-care test (PoCT) meters provide convenient, rapid measures of sUA levels to monitor/adjust therapy. Four commercially available sUA PoCT meters were compared qualitatively (ease of use) and quantitatively (precision/accuracy). METHODS: Liquid chromatography mass spectrometry (LC-MS) analysis was used as a laboratory reference method to compare precision of PoCT devices using blood samples from 20 healthy volunteers. Accuracy was compared against the LC-MS reference method and a laboratory clinical chemistry platform. RESULTS: Qualitatively, EasyTouch(®) GU and UASure were least user-friendly, requiring repeated attempts for accurate use. HumaSens and BeneCheck provided good usability and assay precision. Cross-validation of meter precision with laboratory-based uricase assay gave good correlations between both methods (R(2) = 0.8061 and 0.7605, respectively). CONCLUSION: UA PoCT meters can be important in managing gout, and the characteristics compared herein may enhance successful patient uptake.