Your browser doesn't support javascript.
loading
The influence of varying the number of characters per row on the accuracy and reproducibility of the ETDRS visual acuity chart.
Shamir, Reuben R; Friedman, Yael G; Joskowicz, Leo; Mimouni, Michael; Blumenthal, Eytan Z.
Afiliação
  • Shamir RR; School of Engineering and Computer Science, Hebrew University, Jerusalem, Israel.
  • Friedman YG; Department of Ophthalmology, Rambam Medical Center, P.O.B 9602, Haifa, 31096, Israel.
  • Joskowicz L; School of Engineering and Computer Science, Hebrew University, Jerusalem, Israel.
  • Mimouni M; Department of Ophthalmology, Rambam Medical Center, P.O.B 9602, Haifa, 31096, Israel.
  • Blumenthal EZ; Department of Ophthalmology, Rambam Medical Center, P.O.B 9602, Haifa, 31096, Israel. e_blumenthal@rambam.health.gov.il.
Graefes Arch Clin Exp Ophthalmol ; 254(5): 971-6, 2016 May.
Article em En | MEDLINE | ID: mdl-26743750
ABSTRACT

BACKGROUND:

As part of an effort to improve upon the Snellen chart, we provide a standardized version of the ETDRS chart utilizing five characters in each row. The choice of five characters contradicts the recommended ten characters per row determined by the NAS-NRC, a committee established to provide guidelines for testing visual acuity. We set out to quantify the influence of varying the number of characters per line on the ETDRS chart with respect to the accuracy and reproducibility of visual acuity measurement.

METHODS:

Eleven different ETDRS charts were created, each with a different number of characters appearing in each row. A computer simulation was programmed to run 10,000 virtual patients, each with a unique visual acuity, false-positive and false-negative error value.

RESULTS:

Accuracy and reproducibility were found to roughly correlate with the number of characters present in each row, such that charts with 1, 3, 5, 7, 9, and 11 characters per row provided accuracy of 0.164, 0.094, 0.078, 0.073, 0.071, and 0.070 logMAR, respectively. A non-linear relationship was observed, with little improvement found beyond seven characters per row. In addition, charts with an even number of characters per row provided higher accuracy than their greater-number odd counterparts. In certain instances, accuracy and reproducibility were not well correlated.

CONCLUSIONS:

Increasing the number of characters per row in the ETDRS chart provides a trade-off between accuracy and test duration. An optimized chart layout would take these findings into account, allowing for the use of different chart layouts for clinical versus research settings.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Testes Visuais / Simulação por Computador / Acuidade Visual Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Testes Visuais / Simulação por Computador / Acuidade Visual Idioma: En Ano de publicação: 2016 Tipo de documento: Article