Concentration Monitoring of Plasma Ribavirin: Validation of a Liquid Chromatography-Mass Spectrometric Method and Clinical Sample Collection.
Ther Drug Monit
; 38(1): 50-8, 2016 Feb.
Article
em En
| MEDLINE
| ID: mdl-26766748
BACKGROUND: A liquid chromatography-tandem mass spectrometric (LC-MS/MS) method was developed for routine measurement of ribavirin concentrations in EDTA-anticoagulated plasma. METHODS: After protein precipitation, we used a bridged ethylene hybrid (hydrophilic interaction) chromatography column, 0.1 mmol/L ammonium formate pH 3.0, and a gradient of 85%-96% acetonitrile to achieve baseline separation of ribavirin from isobaric uridine. Quantitation was assured using both primary (m/z 245.3 > 113.0) and secondary transitions (m/z 245.3 > 96.0) of the protonated species. Chromatographic separation and column washing also negated interference from major phospholipid species. RESULTS: There was a linear relationship between concentration and response to 10 mg/L, with a minimum detectable level and a minimum level of quantitation both of 0.1 mg/L. Imprecision within the assay was <10% at 0.1 mg/L and <6% between assays for concentrations >0.4 mg/L. Bias was <4%. In clinical samples (n = 12), there was no difference in ribavirin concentrations obtained by an established liquid chromatographic assay with ultraviolet detection. Ribavirin concentrations were stable in plasma stored at room temperature for 3 days but then decreased significantly on day 7. Plasma concentrations were stable for 15 weeks at -20 °C. Concentrations in plasma separated from whole blood at room temperature fell by a median of 19.4% at 4 hours and then rose substantially (median 251% by 3 days). Dose-normalized ribavirin concentrations reached a steady state after a mean of >6 weeks treatment in 76 patients with hepatitis C. CONCLUSIONS: A hydrophilic interaction liquid chromatography-tandem mass spectrometric method to measure ribavirin in plasma was developed. Samples for ribavirin estimation should be kept at 4 °C, separated within 2 hours of collection and stored at 4 °C before analysis, with long-term storage at -20 °C. This method was applied to a study of the ribavirin therapeutic monitoring in patients with hepatitis C.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Ribavirina
/
Cromatografia Líquida
/
Monitoramento de Medicamentos
/
Espectrometria de Massas em Tandem
Idioma:
En
Ano de publicação:
2016
Tipo de documento:
Article