Comparison of myocardial T1 and T2 values in 3 T with T2* in 1.5 T in patients with iron overload and controls.
Int J Hematol
; 103(5): 530-6, 2016 May.
Article
em En
| MEDLINE
| ID: mdl-26872908
Myocardial iron quantification remains limited to 1.5 T systems with T2* measurement. The present study aimed at comparing myocardial T2* values at 1.5 T to T1 and T2 mapping at 3.0 T in patients with iron overload and healthy controls. A total of 17 normal volunteers and seven patients with a history of myocardial iron overload were prospectively enrolled. Mid-interventricular septum T2*, native T1 and T2 times were quantified on the same day, using a multi-echo gradient-echo sequence at 1.5 T and T1 and T2 mapping sequences at 3.0 T, respectively. Subjects with myocardial iron overload (T2* < 20 ms) in comparison with those without had significantly lower mean myocardial T1 times (868.9 ± 120.2 vs. 1170.3 ± 25.0 ms P = 0.005 respectively) and T2 times (34.9 ± 4.7 vs. 45.1 ± 2.0 ms P = 0.007 respectively). 3 T T1 and T2 times strongly correlated with 1.5 T, T2* times (Pearson's r = 0.95 and 0.91 respectively). T1 and T2 measures presented less variability than T2* in inter- and intra-observer analysis. Native myocardial T1 and T2 times at 3 T correlate closely with T2* times at 1.5 T and may be useful for myocardial iron overload quantification.