Your browser doesn't support javascript.
loading
Comparing Glaucomatous Disc Change Using Stereo Disc Viewing and the MatchedFlicker Software Program in Ophthalmologists-in-Training.
Schaefer, Jamie L; Lukowski, Zachary L; Meyer, Alissa M; Leoncavallo, Anthony J; Greer, Anthony; Martorana, Gina M; Zou, Baiming; Shuster, Jonathan J; Sherwood, Mark B.
Afiliação
  • Schaefer JL; University of Florida, Department of Ophthalmology, Gainesville, Florida; Ross Eye Institute, University at Buffalo, Department of Ophthalmology, Buffalo, New York.
  • Lukowski ZL; University of Florida, Department of Ophthalmology, Gainesville, Florida.
  • Meyer AM; University of Florida, Department of Ophthalmology, Gainesville, Florida.
  • Leoncavallo AJ; University of Florida, Department of Ophthalmology, Gainesville, Florida.
  • Greer A; University of Florida, Department of Ophthalmology, Gainesville, Florida.
  • Martorana GM; University of Florida, Department of Ophthalmology, Gainesville, Florida.
  • Zou B; University of Florida, Department of Ophthalmology, Gainesville, Florida.
  • Shuster JJ; University of Florida, Department of Ophthalmology, Gainesville, Florida.
  • Sherwood MB; University of Florida, Department of Ophthalmology, Gainesville, Florida. Electronic address: sherwood@ufl.edu.
Am J Ophthalmol ; 167: 88-95, 2016 Jul.
Article em En | MEDLINE | ID: mdl-27038890
ABSTRACT

PURPOSE:

To compare the accuracy and speed of using the computerized MatchedFlicker software program (EyeIC Inc, Narberth, Pennsylvania, USA) to evaluate glaucomatous optic disc change against the traditional gold standard of manually examining stereoscopic disc photographs.

DESIGN:

A prospective evaluation of diagnostic technology.

METHODS:

Two resident ophthalmologists and 1 glaucoma fellow at the University of Florida independently evaluated 140 image pairs from 100 glaucomatous/ocular hypertensive patient eyes using a handheld stereo viewer and the MatchedFlicker program. Fifty had progression to glaucoma as determined by the Ocular Hypertension Treatment Study (OHTS) Optic Disc Reading Group and the OHTS Endpoint Committee in the OHTS, and 50 more had photographs taken a few minutes apart, which were negative controls with no progression. Twenty photograph pairs from each group were duplicated to determine reviewer variability. Photographs were examined in alternating blocks of 70 photograph pairs for each method, with the starting viewing method randomized. Reviewer accuracy and time to review for each method were measured.

RESULTS:

Using the handheld stereo viewer, the reviewers correctly identified progression or nonprogression in 76.0% of the slide pairs. Using the MatchedFlicker software, 87.6% were correctly identified (P = .011). Evaluator speed averaged 34.1 seconds per image pair with the stereo viewer vs 24.9 seconds with the MatchedFlicker program (P = .044). Overall, Flicker was significantly more specific but less sensitive than stereo slides. Trainees appeared more reluctant to identify glaucoma progression from slides than from Flicker. For the 2 less experienced trainees Flicker was significantly more accurate.

CONCLUSION:

The MatchedFlicker software had a greater accuracy and was quicker to perform than using a handheld stereoscopic viewer.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Oftalmologia / Disco Óptico / Ensino / Diagnóstico por Imagem / Doenças do Nervo Óptico / Glaucoma de Ângulo Aberto / Técnicas de Diagnóstico Oftalmológico Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Oftalmologia / Disco Óptico / Ensino / Diagnóstico por Imagem / Doenças do Nervo Óptico / Glaucoma de Ângulo Aberto / Técnicas de Diagnóstico Oftalmológico Idioma: En Ano de publicação: 2016 Tipo de documento: Article