Ambient temperature-independent dual-band mid-infrared radiation thermometry.
Appl Opt ; 55(9): 2169-74, 2016 03 20.
Article em En | MEDLINE | ID: mdl-27140549
ABSTRACT
For temperature measurements of targets at low temperatures, dual-band radiation thermometry using mid-infrared detectors has been investigated extensively. However, the accuracy is greatly affected by the reflected ambient radiation and stray radiation, which depend on the ambient temperature. To ensure measurement accuracy, an improved dual-band measurement model is established by considering the reflected ambient radiation and the stray radiation. The effect of ambient temperature fluctuation on temperature measurement is then further analyzed in detail. Experimental results of measuring a gray-body confirm that the proposed method yields high accuracy at varying ambient temperatures. This method provides a practical approach to remove the effect of ambient temperature fluctuations on temperature measurements.

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2016 Tipo de documento: Article