Opening-closing pattern of four pericardial prostheses: results from an in vitro study of leaflet kinematics.
Tasca, Giordano; Fiore, Gianfranco Beniamino; Mangini, Andrea; Romagnoni, Claudia; Gamba, Amando; Redaelli, Alberto; Antona, Carlo; Vismara, Riccardo.
Afiliação
  • Tasca G; Cardiac Surgery Unit, Cardiovascular Department, Ospedale "A. Manzoni" di Lecco, Via Dell'Eremo 9/11, 23900, Lecco, Italy. giordano.tasca67@gmail.com.
  • Fiore GB; Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Milan, Italy. giordano.tasca67@gmail.com.
  • Mangini A; Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Milan, Italy.
  • Romagnoni C; ForCardiolab, Fondazione per la Ricerca in Cardiochirurgia ONLUS, Milan, Italy.
  • Gamba A; Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Milan, Italy.
  • Redaelli A; Cardiovascular Surgery Depatment, "L. Sacco" Hospital, Università degli Studi di Milano, Milan, Italy.
  • Antona C; ForCardiolab, Fondazione per la Ricerca in Cardiochirurgia ONLUS, Milan, Italy.
  • Vismara R; Cardiovascular Surgery Depatment, "L. Sacco" Hospital, Università degli Studi di Milano, Milan, Italy.
J Artif Organs ; 19(4): 350-356, 2016 Dec.
Article em En | MEDLINE | ID: mdl-27230086
ABSTRACT
Pericardial and porcine stented aortic valves have different leaflet kinematics. To study the biomechanics of a prosthesis thoroughly, the in vitro setting is the most appropriate. The aim of our study was to find out whether the prosthesis design in which the pericardial sheet is outside the stent post might influence the opening and closing patterns of the leaflets. Four pericardial prostheses (Magna Ease [MG] 21, Trifecta [TRI] 21, Soprano-Armonia [SA] 20 and Mitroflow [MF] 23) that fitted aortic roots with a native annulus diameter of 2.1 cm were implanted and their leaflet kinematics was studied by a high-speed digital camera. In the opening phase, MG showed the shortest RVOT and the highest RVOVI, with values of 12 ± 2 and 209 ± 17 ms, respectively. The RVOT of MG was significantly shorter than that of MF (p < 0.01), but not than that of TRI (p = 0.286). Both TRI and SA showed similar opening patterns (TRI RVOT of 15 ± 3 ms and RVOVI of 132 ± 25 ms; SA 17 ± 2 ms and 126 ± 19 ms), without statistically significant difference. Conversely, MF showed the slowest profile, with an RVOT of 23 ± 3 ms and an RVOVI of 94 ± 8 ms (Table 1; Fig. 3). The opening/closing profile is not influenced by the position of the pericardial leaflets, but depends on other intrinsic structural characteristics related to the material used for the stent and leaflets. Moreover, the kinematics does not affect the valve performance. Table 1 Kinematics and hydrodynamic results, reported as means and standard deviations, evaluated over the tested heart samples TRI SA MG MF ANOVA TRI versus SA TRI versus MG TRI versus MF SA versus MG SA versus MF MG versus MF p Value p Value p Value p Value p Value p Value p Value ET (ms) 1.0 1.0 1.0 1.0 RVOT (ms) 15 ± 3 17 ± 2 12 ± 2 23 ± 3 <0.01 1.0 0.286 <0.01 0.03 <0.01 <0.01 SVCT (ms) 247 ± 14 231 ± 15 256 ± 26 241 ± 11 0.170 0.463 0.853 0.931 0.213 1.0 1.0 RVCT (ms) 35 ± 19 52 ± 13 32 ± 17 52 ± 4 0.07 0.474 1.0 0.494 0.236 1.0 0.247 TVCT (ms) 283 ± 10 283 ± 19 289 ± 10 293 ± 11 0.584 1.00 1.0 1.0 1.0 1.0 1.0 RVOVI (ms-1) 132 ± 25 126 ± 19 209 ± 17 94 ± 8 <0.01 0.959 <0.01 0.02 <0.01 0.07 <0.01 SVCVI (ms-1) -0.9 ± 0.3 -1.1 ± 0.4 -0.57 ± 0.1 -0.55 ± 0.1 <0.01 1.0 0.353 0.292 0.045 0.04 1.0 RVCVI (ms-1) -16 ± 4 -10 ± 2 -18 ± 6 -10 ± 1 <0.01 0.396 1.0 0.513 0.025 1.0 0.03 Δp (mmHg) 6.7 ± 3.6 10.6 ± 5.5 15.2 ± 7.9 10.7 ± 6.1 <0.01 0.01 <0.01 0.01 0.04 1.0 <0.01 EOA (cm2) 2.2 ± 1.2 1.7 ± 0.9 1.5 ± 0.8 1.7 ± 0.9 <0.01 0.03 <0.01 0.01 0.261 0.617 0.11 El  % 7.3 ± 1 11.9 ± 1 15.4 ± 2 11.8 ± 3 <0.01 <0.01 <0.01 <0.01 0.04 1.00 0.03 CO (L/min) 3.1 ± 0.4 2.8 ± 0.5 3.1 ± 0.3 3.0 ± 0.5 0.534 0.282 0.792 0.702 0.106 0.552 0.559 ET ejection time, RVOT rapid valve-opening time, SVCT slow valve-closing time, RVCT rapid valve-closing time, TVCT total valve-closing time, RVOVI rapid valve-opening velocity index, SVCVI slow valve-closing velocity index, RVCVI rapid valve-closing velocity index, Δp mean pressure drop, EOA effective orifice area, El % energy loss, CO cardiac output.