Breast dose reduction for chest CT by modifying the scanning parameters based on the pre-scan size-specific dose estimate (SSDE).
Kidoh, Masafumi; Utsunomiya, Daisuke; Oda, Seitaro; Nakaura, Takeshi; Funama, Yoshinori; Yuki, Hideaki; Hirata, Kenichiro; Namimoto, Tomohiro; Sakabe, Daisuke; Hatemura, Masahiro; Yamashita, Yasuyuki.
Afiliação
  • Kidoh M; Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan. masafkidoh@yahoo.co.jp.
  • Utsunomiya D; Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan.
  • Oda S; Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan.
  • Nakaura T; Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan.
  • Funama Y; Department of Medical Physics, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan.
  • Yuki H; Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan.
  • Hirata K; Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan.
  • Namimoto T; Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan.
  • Sakabe D; Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan.
  • Hatemura M; Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan.
  • Yamashita Y; Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, 1-1-1, Honjo, Kumamoto, 860-8556, Japan.
Eur Radiol ; 27(6): 2267-2274, 2017 Jun.
Article em En | MEDLINE | ID: mdl-27718079
ABSTRACT

OBJECTIVE:

To investigate the usefulness of modifying scanning parameters based on the size-specific dose estimate (SSDE) for a breast-dose reduction for chest CT. MATERIALS AND

METHODS:

We scanned 26 women with a fixed volume CT dose index (CTDIvol) (15 mGy) and another 26 with a fixed SSDE (15 mGy) protocol (protocol 1 and 2, respectively). In protocol 2, tube current was calculated based on the patient habitus obtained on scout images. We compared the mean breast dose and the inter-patient breast dose variability and performed linear regression analysis of the breast dose and the body mass index (BMI) of the two protocols.

RESULTS:

The mean breast dose was about 35 % lower under protocol 2 than protocol 1 (10.9 mGy vs. 16.8 mGy, p < 0.01). The inter-patient breast dose variability was significantly lower under protocol 2 than 1 (1.2 mGy vs. 2.5 mGy, p < 0.01). We observed a moderate negative correlation between the breast dose and the BMI under protocol 1 (r = 0.43, p < 0.01); there was no significant correlation (r = 0.06, p = 0.35) under protocol 2.

CONCLUSION:

The SSDE-based protocol achieved a reduction in breast dose and in inter-patient breast dose variability. KEY POINTS • CT scan parameters can be modified based on the pre-scan SSDE. • The pre-scan SSDE is useful for a breast dose reduction. • The fixed SSDE protocol reduced individual variations in the breast dose.