Your browser doesn't support javascript.
loading
Performance evaluation of the touchscreen-based Muse™ Auto CD4/CD4% single-platform system for CD4 T cell numeration in absolute number and in percentage using blood samples from children and adult patients living in the Central African Republic.
Mossoro-Kpinde, Christian Diamant; Kouabosso, André; Mboumba Bouassa, Ralph-Sydney; Longo, Jean De Dieu; Kokanzo, Edouard; Féissona, Rosine; Grésenguet, Gérard; Bélec, Laurent.
Afiliação
  • Mossoro-Kpinde CD; Laboratoire National de Biologie Clinique et de Santé Publique, Bangui, Central African Republic. mossoro_kpinde@yahoo.fr.
  • Kouabosso A; Faculté des Sciences de la Santé, Université de Bangui, Bangui, Central African Republic. mossoro_kpinde@yahoo.fr.
  • Mboumba Bouassa RS; Faculté des Sciences de la Santé, Université de Bangui, Bangui, Central African Republic.
  • Longo JD; Centre National de Référence des Infection Sexuellement Transmissibles et de la Thérapie Antirétrovirale, Bangui, Central African Republic.
  • Kokanzo E; Laboratoire de virologie, Hôpital Européen Georges Pompidou, Paris, France.
  • Féissona R; Université Paris Descartes, Paris Sorbonne Cité, Paris, France.
  • Grésenguet G; Faculté des Sciences de la Santé, Université de Bangui, Bangui, Central African Republic.
  • Bélec L; Unité de Recherches et d'Intervention sur les Maladies Sexuellement Transmissibles et le SIDA, Département de Santé Publique, Faculté des Sciences de la Santé de Bangui, Bangui, Central African Republic.
J Transl Med ; 14(1): 326, 2016 11 25.
Article em En | MEDLINE | ID: mdl-27884153
ABSTRACT

BACKGROUND:

The new microcapillary and fluorescence-based EC IVD-qualified Muse™ Auto CD4/CD4% single-platform assay (EMD Millipore Corporation, Merck Life Sciences, KGaA, Darmstadt, Germany) for CD4 T cell numeration in absolute number and in percentage was evaluated using Central African patients' samples compared against the reference EC IVD-qualified BD FACSCount (Becton-Dickinson, USA) flow cytometer.

METHODS:

EDTA-blood samples from 124 adults, 10 adolescents, 13 children and 3 infants were tested in parallel at 2 reference laboratories in Bangui.

RESULTS:

The Musetechnique was highly reproducible, with low intra- and inter-run variabilities less than 15%. CD4 T cell counts of Muse™ and BD FACSCount in absolute number and percentage were highly correlated (r2 = 0.99 and 0.98, respectively). The mean absolute bias between Muse™ and BD FACSCount cells in absolute number and percentage were -5.91 cells/µl (95% CI -20.90 to 9.08) with limits of agreement from -77.50 to 202.40 cells/µl, and +1.69 %CD4 (95% CI ±1.29 to +2.09), respectively. The percentages of outliers outside the limits of agreement were nearly similar in absolute number (8%) and percentage (10%). CD4 T cell counting by Muse™ allowed identifying the majority of individuals with CD4 T cell <200, <350 or <750 cells/µl corresponding to the relevant thresholds of therapeutic care, with sensitivities of 95.5-100% and specificities of 83.9-100%.

CONCLUSIONS:

The Muse™ Auto CD4/CD4% Assay analyzer is a reliable alternative flow cytometer for CD4 T lymphocyte enumeration to be used in routine immunological monitoring according to World Health Organization recommendations in HIV-infected adults as well as children living in resource-constrained settings.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Linfócitos T CD4-Positivos / Contagem de Linfócito CD4 Idioma: En Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Linfócitos T CD4-Positivos / Contagem de Linfócito CD4 Idioma: En Ano de publicação: 2016 Tipo de documento: Article