Your browser doesn't support javascript.
loading
Unilateral Spinous Process Noncovering Hook Type Patient-specific Drill Template for Thoracic Pedicle Screw Fixation: A Pilot Clinical Trial and Template Classification.
Kim, Sang B; Won, Yougun; Yoo, Hyun Jin; Sin, Lee J; Rhee, John M; Lee, Sang W; Lee, Gi-Soo.
Afiliação
  • Kim SB; College of Medicine, Konyang University, Boston, MA.
  • Won Y; College of Medicine, Konyang University, Boston, MA.
  • Yoo HJ; College of Medicine, Konyang University, Boston, MA.
  • Sin LJ; College of Medicine, Konyang University, Boston, MA.
  • Rhee JM; Department of Orthopedic Surgery, Emory University, Atlanta, GA.
  • Lee SW; School of Biomedical Engineering, Konyang University, Boston, MA.
  • Lee GS; College of Medicine, Konyang University, Boston, MA.
Spine (Phila Pa 1976) ; 42(18): E1050-E1057, 2017 Sep 15.
Article em En | MEDLINE | ID: mdl-28092341
ABSTRACT
STUDY

DESIGN:

Prospective clinical pilot study.

OBJECTIVE:

To confirm the accuracy of thoracic pedicle screw placement by using the unilateral spinous noncovering hook type patient-specific drill template (PSDT) made through rapid prototyping (RP) and to analyze previously reported PSDT designs and their characteristics. SUMMARY OF BACKGROUND DATA Pedicle screw fixation is the most common form of the posterior instrumentation of the thoracic and lumbar spine. Various techniques have been introduced to improve pedicle screw placement. Among them PSDT with a preplanned trajectory has been considered a promising solution; however, we don't have consensus on proper character of the template.

METHODS:

Preoperative spiral three-dimensional (3D) computed tomography (CT) was performed on the thoracic spine. The images were stored in DICOM format and transferred to a workstation running MIMICS 17.0 software to generate a 3D reconstruction template for the desired thoracic vertebra. The accurate trajectory and screw diameter and length were calculated with UG Imageware 12.1. The guide template was sterilized and used intraoperatively to assist with the placement of thoracic pedicle screws. After all pedicle trajectory screws had been inserted. We reviewed 12 previous reports and classified them according to the shape and system of PSDT that met the inclusion criteria of the review.

RESULTS:

Ten screws were placed by using the PSDT without violating the single laminar cortex. There was no violation of the spinal canal or the cortex of pedicle on postoperative CT scans. The results of 13 PSDT types included in the current study suggested that there is no significant difference in accuracy between the PSDTs.

CONCLUSION:

The unilateral spinous process noncovering hook type PSDT made through RP provided an accurate trajectory for the thoracic vertebra, and the classification of PSDT in this study could be helpful for further studies. LEVEL OF EVIDENCE 5.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Vértebras Torácicas / Procedimentos Ortopédicos / Cirurgia Assistida por Computador / Parafusos Pediculares Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Vértebras Torácicas / Procedimentos Ortopédicos / Cirurgia Assistida por Computador / Parafusos Pediculares Idioma: En Ano de publicação: 2017 Tipo de documento: Article