Analysis of stress-only imaging, comparing upright and supine CZT camera acquisition to conventional gamma camera images with and without attenuation correction, with coronary angiography as a reference.
Jameria, Zenith A; Abdallah, Mouhamad; Fernandez-Ulloa, Mariano; O'Donnell, Robert; Dwivedi, Alok K; Washburn, Erica; Khan, Naseer; Khaleghi, Mahyar; Kalakota, Nischelle; Gerson, Myron C.
Afiliação
  • Jameria ZA; Division of Cardiology, Department of Internal Medicine, University of Illinois at Chicago, Chicago, IL, USA.
  • Abdallah M; Division of Cardiology, Department of Internal Medicine, University of Cincinnati Medical Center, Cincinnati, OH, USA.
  • Fernandez-Ulloa M; Division of Nuclear Medicine, Department of Radiology, University of Cincinnati Medical Center, Cincinnati, OH, USA.
  • O'Donnell R; Division of Cardiology, Department of Internal Medicine, University of Cincinnati Medical Center, Cincinnati, OH, USA.
  • Dwivedi AK; Division of Biostatistics and Epidemiology, Department of Biomedical Sciences, Texas Tech University Health Sciences Center, El Paso, TX, USA.
  • Washburn E; Division of Nuclear Medicine, Department of Radiology, University of Cincinnati Medical Center, Cincinnati, OH, USA.
  • Khan N; Division of Cardiology, Department of Internal Medicine, University of Cincinnati Medical Center, Cincinnati, OH, USA.
  • Khaleghi M; Division of Cardiology, Department of Internal Medicine, University of Cincinnati Medical Center, Cincinnati, OH, USA.
  • Kalakota N; University of Cincinnati College of Medicine, Cincinnati, USA.
  • Gerson MC; Division of Cardiology, Department of Internal Medicine, University of Cincinnati Medical Center, Cincinnati, OH, USA. myron.gerson@uc.edu.
J Nucl Cardiol ; 25(2): 540-549, 2018 04.
Article em En | MEDLINE | ID: mdl-28108979
ABSTRACT

BACKGROUND:

Diagnostic performance of stress-only imaging using a Cadmium-Zinc-Telluride (CZT) camera has not been directly compared in the same patients to stress-only attenuation-corrected conventional Anger camera images.

METHODS:

112 subjects with correlative coronary angiographic data and 40 subjects with <5% pre-test likelihood of coronary disease completed attenuation-corrected stress-only images on a conventional Anger camera and uncorrected upright and supine stress images on a CZT camera. Two readers provided independent, blinded interpretations of stress-only images.

RESULTS:

Upright and supine stress-only CZT images and attenuation-corrected Anger camera images provided similar positive (reader 1/reader 2, 50.0%/44.1% vs 46.4%/51.9%) and negative (66.7%/64.0% vs 67.9%/67.7%) predictive values (all P = NS) for obstructive coronary artery disease; however, the sensitivity was higher (81.3% vs 58.3%, P = .05), specificity lower (29.7% vs 50.0%, P = .005), and normalcy rate lower (87.5% vs 100%, P = .025) with attenuation-corrected Anger camera images for the first reader with no significant differences between cameras for the second reader.

CONCLUSIONS:

Stress-only upright and supine CZT imaging was non-inferior statistically to attenuation-corrected stress-only Anger camera imaging. Nevertheless, stress-only CZT imaging may be associated with reduced diagnostic sensitivity for some readers compared to attenuation-corrected Anger camera images, which may be less acceptable clinically compared to stress plus rest images.