A Method for the Automatic Exposure Control in Pediatric Abdominal CT: Application to the Standard Deviation Value and Tube Current Methods by Using Patient's Age and Body Size.
Furuya, Ken; Akiyama, Shinji; Nambu, Atushi; Suzuki, Yutaka; Hasebe, Yuusuke.
Afiliação
  • Furuya K; Department of Radiology, Municipal Kofu Hospital.
Article em Ja | MEDLINE | ID: mdl-28111396
We aimed to apply the pediatric abdominal CT protocol of Donnelly et al. in the United States to the pediatric abdominal CT-AEC. Examining CT images of 100 children, we found that the sectional area of the hepatic portal region (y) was strongly correlated with the body weight (x) as follows: y=7.14x + 84.39 (correlation coefficient=0.9574). We scanned an elliptical cone phantom that simulates the human body using a pediatric abdominal CT scanning method of Donnelly et al. in, and measured SD values. We further scanned the same phantom under the settings for adult CT-AEC scan and obtained the relationship between the sectional areas (y) and the SD values. Using these results, we obtained the following preset noise factors for CT-AEC at each body weight range: 6.90 at 4.5-8.9 kg, 8.40 at 9.0-17.9 kg, 8.68 at 18.0-26.9 kg, 9.89 at 27.0-35.9 kg, 12.22 at 36.0-45.0 kg, 13.52 at 45.1-70.0 kg, 15.29 at more than 70 kg. From the relation between age, weight and the distance of liver and tuber ischiadicum of 500 children, we obtained the CTDIvol values and DLP values under the scanning protocol of Donnelly et al. Almost all of DRL from these values turned out to be smaller than the DRL data of IAEA and various countries. Thus, by setting the maximum current values of CT-AEC to be the Donnelly et al.'s age-wise current values, and using our weight-wise noise factors, we think we can perform pediatric abdominal CT-AEC scans that are consistent with the same radiation safety and the image quality as those proposed by Donnelly et al.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Abdome Idioma: Ja Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Abdome Idioma: Ja Ano de publicação: 2017 Tipo de documento: Article