SU-E-T-122: Dose Response Analysis of Radiochromic Films in Regions of Low Dose Using Separation Color Components.
Med Phys
; 39(6Part11): 3731, 2012 Jun.
Article
em En
| MEDLINE
| ID: mdl-28517164
PURPOSE: To evaluate the dose response of EBT2 films in regions of low dose using the decomposition of the image's color channels (RGB, Red, Green and Blue). METHODS: Doses ranging from 1 Gy to 60 Gy were used to calibrate the dose response of Gafchromic ® EBT2 films irradiated in 6MV photons beams. Segments of film with dimensions of 8.5 cm × 8.5 cm were used. Another segment of film with dimensions 8.5 cm × 20.25 cm was also irradiated with a maximum dose of 4Gy to determine the percentage depth dose (PDD). The films were digitized by a LaserJet M1132 MFP - HP ® scanner in standard resolution of 150dpi and analyzed by a routine created in MatLab to convert the image to gray levels as well as assess the desired color components from the image. RESULTS: The green component presented the higher sensitivity (17.8 a.u./Gy) when the separated color channels and the shades of gray analysis are compared. The red component presented the highest signal to noise ratio in the low dose range (63% at 1Gy). The blue component presented low sensitivity (0.66 a.u./Gy) in the entire dose range. A linear fitting (r=0.998) was used to the green and gray components until a dose of 4 Gy. The red component presented a non-linear behavior in the entire dose range. The useful dose range found was from 1 Gy to 15 Gy. The maximum differences between the reference PDD, measured with ionization chamber in a water phantom, and the PDDs determined with film were 6%, 9% and 14% for the green, gray and red components, respectively. CONCLUSIONS: This work results show that the use of radiochromic films on planning verification procedures in low dose ranges can be benefit from the analysis of the image's separated color components.
Texto completo:
1
Base de dados:
MEDLINE
Idioma:
En
Ano de publicação:
2012
Tipo de documento:
Article