Quantitative detection of glucose level based on radiofrequency patch biosensor combined with volume-fixed structures.
Qiang, Tian; Wang, Cong; Kim, Nam-Young.
Afiliação
  • Qiang T; Department of Electronic Engineering, Kwangwoon University, Seoul, Republic of Korea.
  • Wang C; Department of Microwave Engineering, Harbin Institute of Technology, Harbin, China. Electronic address: kevinwang@hit.edu.cn.
  • Kim NY; Department of Electronic Engineering, Kwangwoon University, Seoul, Republic of Korea. Electronic address: nykim@kw.ac.kr.
Biosens Bioelectron ; 98: 357-363, 2017 Dec 15.
Article em En | MEDLINE | ID: mdl-28704784
ABSTRACT
A concept for characterizing a radiofrequency (RF) patch biosensor combined with volume-fixed structures is presented for timely monitoring of an individual's glucose levels based on frequency variation. Two types of patch biosensors-separately integrated with a backside slot (0.53µL) and a front-side tank (0.70µL) structure-were developed to achieve precise and efficient detection while excluding the effects of interference due to the liquidity, shape, and thickness of the tested glucose sample. A glucose test analyte at different concentrations (50-600mg/dL) was dropped into the volume-fixed structures. It fully interacted with the RF patch electromagnetic field, effectively and sensitively changing the resonance frequency and magnitude of the reflection coefficient. Measurement results based on the resonance frequency showed high sensitivity up to 1.13MHz and 1.97MHz per mg/dL, and low detection limits of 26.54mg/dL and 15.22mg/dL, for the two types of patch biosensors, respectively, as well as a short response time of less than 1s. Excellent reusability of the proposed biosensors was verified through three sets of measurements for each individual glucose sample. Regression analysis revealed a good linear correlation between glucose concentrations and the resonance frequency shift. Moreover, to facilitate a multi-parameter-sensitive detection of glucose, the magnitude of the reflection coefficient was also tested, and it showed a good linear correlation with the glucose concentration. Thus, the proposed approach can be adopted for distinguishing glucose solution levels, and it is a potential candidate for early-stage detection of glucose levels in diabetes patients.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Glicemia / Técnicas Biossensoriais / Diabetes Mellitus / Glucose Idioma: En Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Glicemia / Técnicas Biossensoriais / Diabetes Mellitus / Glucose Idioma: En Ano de publicação: 2017 Tipo de documento: Article