Tissue oxygenation mapping by combined chemical shift and T1 magnetic resonance imaging.
Franconi, Florence; Lemaire, Laurent; Saint-Jalmes, Hervé; Saulnier, Patrick.
Afiliação
  • Franconi F; PRISM Plate-forme de recherche en imagerie et spectroscopie multi-modales, PRISM-Icat, Angers et PRISM-Biosit CNRS UMS 3480, INSERM UMS 018, Rennes, UBL Universite Bretagne, Loire, France.
  • Lemaire L; Micro & Nanomédecines Translationelles-MINT, UNIV Angers, INSERM U1066, CNRS UMR 6021, UBL Universite Bretagne Loire, Angers, France.
  • Saint-Jalmes H; PRISM Plate-forme de recherche en imagerie et spectroscopie multi-modales, PRISM-Icat, Angers et PRISM-Biosit CNRS UMS 3480, INSERM UMS 018, Rennes, UBL Universite Bretagne, Loire, France.
  • Saulnier P; Micro & Nanomédecines Translationelles-MINT, UNIV Angers, INSERM U1066, CNRS UMR 6021, UBL Universite Bretagne Loire, Angers, France.
Magn Reson Med ; 79(4): 1981-1991, 2018 04.
Article em En | MEDLINE | ID: mdl-28833514
ABSTRACT

PURPOSE:

To propose a method for determining tissue oxygenation via the measurement of fat T1 . The method is based on a 2D fat/water chemical shift-encoded and T1 -weighted acquisition. THEORY AND

METHODS:

A 2D data set was acquired with a fast spin echo sequence with several echo asymmetries and repetition times, wherein one dimension is related to the fat/water phase modulation and the other to the T1 saturation recovery. A joint magnitude-based process of phase modulation and T1 evolution allowed for the collection of the fat fraction and T1 maps with resolved fat or water dominance ambiguity while avoiding the phased error problem.

RESULTS:

In vitro imaging allowed for the attribution of fat content for different water/oil emulsions that demonstrated longitudinal relaxation rate (R1 ) sensitivity to the oxygenated emulsion environment. The fat R1 values were subsequently compared to reference values, which were measured using low receiver bandwidth acquisition to enhance water and fat signal separations. In vivo feasibility of tissue oxygenation assessment was demonstrated by investigating interscapular brown adipose tissue modifications during an air/carbogen challenge in rats.

CONCLUSION:

The proposed method offers a precise and robust estimate of tissue oxygenation illustrated by the method's ability to detect-brown adipose tissue oxygenation modifications. Magn Reson Med 791981-1991, 2018. © 2017 International Society for Magnetic Resonance in Medicine.