Characterizing and eliminating errors in enhancement and subtraction artifacts in dynamic contrast-enhanced breast MRI: Chemical shift artifact of the third kind.
Derakhshan, Jamal J; McDonald, Elizabeth S; Siegelman, Evan S; Schnall, Mitchell D; Wehrli, Felix W.
Afiliação
  • Derakhshan JJ; Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • McDonald ES; Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Siegelman ES; Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Schnall MD; Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Wehrli FW; Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
Magn Reson Med ; 79(4): 2277-2289, 2018 04.
Article em En | MEDLINE | ID: mdl-28840613
ABSTRACT

PURPOSE:

To characterize errors in enhancement in breast dynamic contrast-enhanced (DCE) MRI studies as a function of echo time and determine the source of dark band artifacts in clinical subtraction images.

METHODS:

Computer simulations, oil and water substitute (methylene chloride), as well as an American College of Radiology quality control phantom were tested. Routine clinical DCE breast MRI study was bracketed with (accelerated) in-phase DCE acquisitions in five patients.

RESULTS:

Simulation results demonstrated up to -160% suppression of the expected enhancement caused by differential enhancement of fat and water. Two-dimensional gradient-recalled echo and fat-suppressed 3D GRE phantom imaging confirmed the simulation results and showed that fat suppression does not eliminate the artifact. In vivo in-phase DCE images showed increased enhancement consistent with predictions and also confirmed increased spatial blurring on in-phase 3D gradient-recalled echo images. Combined multi-dimensional partial Fourier and parallel imaging provided a time-equivalent in-phase DCE MRI acquisition.

CONCLUSION:

Errors in expected enhancement occur in DCE breast MRI subtraction images because of differential enhancement of fat and water and incomplete fat signal suppression. These errors can lead to artificial suppression of enhancement as well as dark band artifacts on subtraction images. These artifacts can be eliminated with a time-equivalent in-phase fat-suppressed 3D gradient-recalled echo sequence. Understanding chemical shift artifact of the third kind, a unique artifact of artificial enhancement suppression in the presence of intravoxel fat and water signal, will aid DCE breast MRI image interpretation. In-phase acquisitions (combined with simultaneous minimum echo time or opposed-phase echoes) may facilitate qualitative, quantitative and longitudinal analysis of contrast enhancement. Magn Reson Med 792277-2289, 2018. © 2017 International Society for Magnetic Resonance in Medicine.