Your browser doesn't support javascript.
loading
Diagnostic yield of asymptomatic arrhythmias detected by mobile cardiac outpatient telemetry and autotrigger looping event cardiac monitors.
Derkac, Wayne M; Finkelmeier, Jeffrey R; Horgan, Daniel J; Hutchinson, Mathew D.
Afiliação
  • Derkac WM; BioTelemetry, Malvern, PA, USA.
  • Finkelmeier JR; BioTelemetry, Malvern, PA, USA.
  • Horgan DJ; BioTelemetry, Malvern, PA, USA.
  • Hutchinson MD; Division of Cardiovascular Medicine, Sarver Heart Center, University of Arizona College of Medicine Tucson, Tucson, AZ, USA.
J Cardiovasc Electrophysiol ; 28(12): 1475-1478, 2017 Dec.
Article em En | MEDLINE | ID: mdl-28940881
ABSTRACT

INTRODUCTION:

Asymptomatic arrhythmias can have important therapeutic implications in certain patient populations, for example, atrial fibrillation in patients with prior ischemic stroke. We sought to compare the diagnostic yield of two commercially available monitoring systems with automated arrhythmia detection algorithms.

METHODS:

We queried a large, proprietary database containing rhythm data for patients receiving ambulatory EKG monitoring (BioTelemetry, Malvern, PA, USA). We compared all patients prescribed mobile cardiac outpatient telemetry (MCOT™, Braemar Manufacturing, LLC, Eagan, MN, USA) versus autotrigger looping event recorder (AT-LER) devices over a consecutive 8-month period. Data from both device types were analyzed for diagnostic yields in detecting asymptomatic (device-triggered) arrhythmias consisting of atrial fibrillation (of any detected duration), bradycardia (ventricular rate ≤ 40 bpm), ventricular pause (≥ 3 seconds), supraventricular tachycardia (≥ 6 consecutive supraventricular beats), and ventricular tachycardia (≥ 4 consecutive premature ventricular contractions). The mean time to first diagnosis of each arrhythmia for each device was determined. Physician-designated diagnostic codes for patients prescribed each device were also determined from the database.

RESULTS:

The MCOT™ device had significantly higher diagnostic yields of all evaluated asymptomatic arrhythmias than the AT-LER. The MCOT™ device also produced an earlier mean time to diagnosis for all evaluated asymptomatic arrhythmias. These findings were noted despite a shorter average prescription length for MCOT™ monitored patients.

CONCLUSIONS:

In patients with conventional diagnostic monitoring indications, MCOT™ had significantly higher diagnostic yields for five asymptomatic arrhythmias compared to the AT-LER.