Improving surgical results in complex nerve anatomy during implantation of selective upper airway stimulation.
Zhu, Zhaojun; Hofauer, Benedikt; Heiser, Clemens.
Afiliação
  • Zhu Z; Department of Otorhinolaryngology, Head and Neck Surgery, Klinikum rechts der Isar, Munich, Germany. Electronic address: zhaojun.zhu@mri.tum.de.
  • Hofauer B; Department of Otorhinolaryngology, Head and Neck Surgery, Klinikum rechts der Isar, Munich, Germany.
  • Heiser C; Department of Otorhinolaryngology, Head and Neck Surgery, Klinikum rechts der Isar, Munich, Germany.
Auris Nasus Larynx ; 45(3): 653-656, 2018 Jun.
Article em En | MEDLINE | ID: mdl-29153261
The following report presents a case of two late embedded hypoglossus branches during implantation of an upper airway stimulation device that caused a mixed activation of the tongue when included in the stimulation cuff. In the end, correct cuff placement could be achieved by careful examination of the hypoglossal nerve anatomy, precise nerve dissection, tongue motion analysis and intraoperative nerve monitoring.