Development of a disposable force-sensing glove for clinicians and demonstration of its force measurements on patients during rehabilitation following anterior cruciate ligament reconstruction surgery.
J Biomed Opt
; 22(12): 1-14, 2017 Dec.
Article
em En
| MEDLINE
| ID: mdl-29264895
ABSTRACT
For many clinicians, their effectiveness is dependent on the magnitude of forces they manually apply to their patients. However, current state-of-the-art care strategies lack quantitative feedback, making it difficult to provide consistent care over time and among multiple clinicians. To provide real-time quantitative feedback to clinicians, we have developed a disposable glove with a force sensor embedded in the fingertips or palm. The sensor is based on the fiber-optic bendloss effect whereby light intensity from an infrared source is attenuated as the fiber is bent between a series of corrugated teeth. The sensor fabricated has a very low profile (10×7×1 mm) and has demonstrated high sensitivity, accuracy, range, and durability. Force feedback up to 90 N with an average force threshold at 0.19 N and average sensor resolution at 0.05 N has been demonstrated. A preliminary clinical study has also been conducted with anterior cruciate ligament reconstruction patients who show significant range of motion improvement when treated with the force-sensing glove.