Accuracy and precision of patient positioning for pelvic MR-only radiation therapy using digitally reconstructed radiographs.
Kemppainen, R; Vaara, T; Joensuu, T; Kiljunen, T.
Afiliação
  • Kemppainen R; Philips MR Therapy, Äyritie 4, FI-01510, Vantaa, Finland. Department of Neuroscience and Biomedical Engineering, Aalto University School of Science, Rakentajanaukio 2 C, FI-02150 Espoo, Finland.
Phys Med Biol ; 63(5): 055009, 2018 03 02.
Article em En | MEDLINE | ID: mdl-29405121
ABSTRACT
BACKGROUND AND

PURPOSE:

Magnetic resonance imaging (MRI) has in recent years emerged as an imaging modality to drive precise contouring of targets and organs at risk in external beam radiation therapy. Moreover, recent advances in MRI enable treatment of cancer without computed tomography (CT) simulation. A commercially available MR-only solution, MRCAT, offers a single-modality approach that provides density information for dose calculation and generation of positioning reference images. We evaluated the accuracy of patient positioning based on MRCAT digitally reconstructed radiographs (DRRs) by comparing to standard CT based workflow. MATERIALS AND

METHODS:

Twenty consecutive prostate cancer patients being treated with external beam radiation therapy were included in the study. DRRs were generated for each patient based on the planning CT and MRCAT. The accuracy assessment was performed by manually registering the DRR images to planar kV setup images using bony landmarks. A Bayesian linear mixed effects model was used to separate systematic and random components (inter- and intra-observer variation) in the assessment. In addition, method agreement was assessed using a Bland-Altman analysis.

RESULTS:

The systematic difference between MRCAT and CT based patient positioning, averaged over the study population, were found to be (mean [95% CI]) -0.49 [-0.85 to -0.13] mm, 0.11 [-0.33 to +0.57] mm and -0.05 [-0.23 to +0.36] mm in vertical, longitudinal and lateral directions, respectively. The increases in total random uncertainty were estimated to be below 0.5 mm for all directions, when using MR-only workflow instead of CT.

CONCLUSIONS:

The MRCAT pseudo-CT method provides clinically acceptable accuracy and precision for patient positioning for pelvic radiation therapy based on planar DRR images. Furthermore, due to the reduction of geometric uncertainty, compared to dual-modality workflow, the approach is likely to improve the total geometric accuracy of pelvic radiation therapy.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pelve / Neoplasias da Próstata / Planejamento da Radioterapia Assistida por Computador / Imageamento por Ressonância Magnética / Posicionamento do Paciente / Radioterapia Guiada por Imagem Idioma: En Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pelve / Neoplasias da Próstata / Planejamento da Radioterapia Assistida por Computador / Imageamento por Ressonância Magnética / Posicionamento do Paciente / Radioterapia Guiada por Imagem Idioma: En Ano de publicação: 2018 Tipo de documento: Article