Maximizing the fixation strength of modular components by impaction without tissue damage.
Krull, A; Morlock, M M; Bishop, N E.
Afiliação
  • Krull A; Institute of Biomechanics, TUHH - Hamburg University of Technology, Denickestrasse 15, 21073 Hamburg, Germany.
  • Morlock MM; Institute of Biomechanics, TUHH - Hamburg University of Technology, Denickestrasse 15, 21073 Hamburg, Germany.
  • Bishop NE; Faculty of Life Sciences, HAW Hamburg University of Applied Sciences, Ulmenliet 20, 21033 Hamburg, Germany.
Bone Joint Res ; 7(2): 196-204, 2018 Feb.
Article em En | MEDLINE | ID: mdl-29682286
OBJECTIVES: Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of this study was to determine the effects of increasing the impaction energy and the stiffness of the impactor tool on the stability of the taper junction and on the forces transmitted through the patient's surrounding tissues. METHODS: A commercially available impaction tool was modified to assemble components in the laboratory using impactor tips with varying stiffness at different applied energy levels. Springs were mounted below the modular components to represent the patient. The pull-off force of the head from the stem was measured to assess stability, and the displacement of the springs was measured to assess the force transmitted to the patient's tissues. RESULTS: The pull-off force of the head increased as the stiffness of the impactor tip increased but without increasing the force transmitted through the springs (patient). Increasing the impaction energy increased the pull-off force but also increased the force transmitted through the springs. CONCLUSIONS: To limit wear and corrosion, manufacturers should maximize the stiffness of the impactor tool but without damaging the surface of the head. This strategy will maximize the stability of the head on the stem for a given applied energy, without influencing the force transmitted through the patient's tissues. Current impactor designs already appear to approach this limit. Increasing the applied energy (which is dependent on the mass of the hammer and square of the contact speed) increases the stability of the modular connection but proportionally increases the force transmitted through the patient's tissues, as well as to the surface of the head, and should be restricted to safe levels.Cite this article: A. Krull, M. M. Morlock, N. E. Bishop. Maximizing the fixation strength of modular components by impaction without tissue damage. Bone Joint Res 2018;7:196-204. DOI: 10.1302/2046-3758.72.BJR-2017-0078.R2.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2018 Tipo de documento: Article