[The Simulation Experiment of Control System for Left Ventricular Assist Device].
Zhongguo Yi Liao Qi Xie Za Zhi
; 40(2): 143-6, 2016 Mar.
Article
em Zh
| MEDLINE
| ID: mdl-29763537
ABSTRACT
Objective:
To test the accuracy, the reliability and the stability of flow-controlling method of a new percutaneous left ventricular assist device.Methods:
With the AMEsim to set up a simulation model, analysis the variations of motor speed, flow and pressure by adjusting the size of valve. and we test the improved-theoretical model with outside experiment.Results:
In the simulation experiment, with the variations of pressure from A, the system can regulate the motor speed to maintain the pump outlet pressure 1.1×105 Pa~1.2×105 Pa and control the flow. With the B valve size changes, the flow-controlling by the system identify and change the motor speed, to maintain the pump inlet pressure from-0.25×105 Pa to-0.10×105 Pa. In outside experiment, with the variations of the size of valve A and valve B, the pump can change the voltage by self-regulation, maintain the entrance and exit of pressure is relatively stable and control the flow.Conclusions:
It is a fast-responding and better robust way, by adding the pressure sensor to adjust the speed of motor, to control the flow accurately.