Your browser doesn't support javascript.
loading
Effect of a Thin Fluid Layer on Surface Wave Speed Measurements: A Lung Phantom Study.
Zhou, Jinling; Zhang, Xiaoming.
Afiliação
  • Zhou J; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Zhang X; Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
J Ultrasound Med ; 38(5): 1361-1365, 2019 May.
Article em En | MEDLINE | ID: mdl-30208217
ABSTRACT
Lung ultrasound (US) surface wave elastography (SWE) is a novel technique that measures superficial lung tissue elastic properties. A thin pleural fluid layer covers a lung, but its effect on lung measurements in SWE is unknown. We modeled a lung and pleural fluid with sponges and a thin layer of US transmission gel. Sponge surface wave speeds measured from SWE were compared for sponges without and with the thin US gel layer at 3 wave excitation frequencies. The comparison showed that the sponge surface wave speed measurements were not affected by the thin gel layer.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pleura / Líquidos Corporais / Técnicas de Imagem por Elasticidade / Pulmão Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pleura / Líquidos Corporais / Técnicas de Imagem por Elasticidade / Pulmão Idioma: En Ano de publicação: 2019 Tipo de documento: Article