[The First Step in the Optimization of Radiation Protection of Patients in Cerebral Angiography: Investigate the Possibility of Constructing the Diagnostic Reference Level by Imaging Objective/Disease Group Using Display Value of the Blood Vessel Imaging Apparatus].
Hitomi, Go; Matsubara, Shunji; Moritake, Takashi; Sun, Lue; Mura, Masakatsu; Matsumoto, Hiroki; Kusachi, Fumiko; Matsumaru, Yuji; Uno, Masaaki.
Afiliação
  • Hitomi G; Department of Radiological Technology, Kawasaki Medical School Hospital.
  • Matsubara S; Department of Neurosurgery, Kawasaki Medical School Hospital.
  • Moritake T; Department of Radiological Health Science, Institute of Industrial Ecological Sciences, University of Occupational and Environmental Health, Japan.
  • Sun L; Department of Radiation Biology, Faculty of Medicine, University of Tsukuba.
  • Mura M; Department of Radiological Technology, Kawasaki Medical School Hospital.
  • Matsumoto H; Department of Radiological Technology, Kawasaki Medical School Hospital.
  • Kusachi F; Department of Radiological Technology, Kawasaki Medical School Hospital.
  • Matsumaru Y; Division of Stroke Prevention and Treatment, Department of Neurosurgery, Faculty of Medicine, University of Tsukuba.
  • Uno M; Department of Neurosurgery, Kawasaki Medical School Hospital.
Article em Ja | MEDLINE | ID: mdl-30890674
To optimize the radiation protection of patients, we investigated the possibility of constructing the diagnostic reference levels (DRLs) by imaging objective/disease group using display value of the blood vessel imaging apparatus (air kerma-area product: PKA, air kerma at the patient entrance reference point: Ka, r) in cerebral angiography. We used PKA and Ka, r recorded during surgery of 997 patients at our hospital, and classified them according to the purpose of imaging (diagnostic cerebral angiography or neuro interventional radiology) and disease group. Neuro interventional radiology (PKA: 268±155 Gy・cm2, Ka, r: 2420±1462 mGy) was significantly higher than that of diagnostic cerebral angiography (PKA: 161±70 Gy・cm2, Ka, r: 1112±485 mGy), (Mann-Whitney test, P<0.01). Significant difference was found between PKA and Ka, r for imaging purpose and disease group (Kruskal-Wallis test, P<0.05). It is highly probable that the DRL for cerebral angiography can be constructed by imaging purpose/disease group using display value (PKA, Ka, r) of the blood vessel imaging apparatus.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Angiografia Cerebral Idioma: Ja Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Angiografia Cerebral Idioma: Ja Ano de publicação: 2019 Tipo de documento: Article