CT diagnostic reference levels: are they appropriately computed?
Vanaudenhove, Thibault; Van Muylem, Alain; Howarth, Nigel; Gevenois, Pierre Alain; Tack, Denis.
Afiliação
  • Vanaudenhove T; Federal Agency for Nuclear Control (FANC), Rue Ravenstein 36, 1000, Brussels, Belgium.
  • Van Muylem A; Department of Chest Medicine, Hôpital Erasme, Route de Lennik 808, 1070, Brussels, Belgium.
  • Howarth N; Department of Radiology, Clinique des Grangettes, 7 Chemin des Grangettes, 1224, Chêne-Bougeries, Switzerland.
  • Gevenois PA; Department of Radiology, Hôpital Erasme, Route de Lennik 808, 1070, Brussels, Belgium.
  • Tack D; Department of Radiology, Epicura, Clinique Louis Caty, 136 Rue Louis Caty, 7331, Baudour, Belgium. denis.tack@skynet.be.
Eur Radiol ; 29(10): 5264-5271, 2019 Oct.
Article em En | MEDLINE | ID: mdl-30963277
ABSTRACT

OBJECTIVES:

To estimate the variability of CT diagnostic reference levels (DRLs) according to the methods used for computing collected data.

METHODS:

Dose-length products (DLP) were collected by our national nuclear control agency from the 250 devices installed in 140 medical centers in the country. In 2015, the number of head, thorax, abdomen, and lumbar spine examinations collected in these centers ranged from approximately 20,000 to 42,000. The impact on DRLs of the number of devices considered, as well as the differences in descriptive statistics (mean vs. median DLP) or methods of pooling DLP data (all devices vs. all patients), was investigated. Variability in DRLs was investigated using a bootstrapping method as a function of the numbers of devices and examinations per device.

RESULTS:

As expected, DRLs derived from means were higher than those from medians, with substantial differences between device- and patient-related DRLs. Depending on the numbers of devices and DLP data per device, the variability ranged from 10 to 40% but was stabilized at a level of 10-20% if the number of devices was higher than 50 to 60, regardless of the number of DLP data per device.

CONCLUSION:

Number of devices and of DLP data per device, descriptive statistics, and pooling data influence DRLs. As differences in methods of computing survey data can artificially influence DRLs, harmonization among national authorities should be recommended. KEY POINTS • Due to CT dose variability, that of DRLs is at least of 10%. • DRLs derived from medians are lower than from means and differ from those obtained by pooling all patient data. • Fifty to 60 devices should be sufficient for estimating national DRLs, regardless of the number of data collected per device.