Quantitation of trace amorphous solifenacin succinate in pharmaceutical formulations by transmission Raman spectroscopy.
Terada, Hirohito; Hattori, Yusuke; Sasaki, Tetsuo; Otsuka, Makoto.
Afiliação
  • Terada H; Musashino University, Graduate School of Pharmaceutical Sciences, 1-1-20, Shin-machi, Nishi-Tokyo City, Tokyo 202-8585, Japan; Pharmaceutical Research Center, OHARA Pharmaceutical Co., Ltd., 121-15, Toriino, Koka-cho, Koka-shi, Shiga 520-3403, Japan.
  • Hattori Y; Musashino University, Graduate School of Pharmaceutical Sciences, 1-1-20, Shin-machi, Nishi-Tokyo City, Tokyo 202-8585, Japan.
  • Sasaki T; Shizuoka University, Graduate School of Medical of Photonics, 3-5-1 Johoku, Naka-ku, Hamamatsu, Shizuoka 432-8011, Japan.
  • Otsuka M; Musashino University, Graduate School of Pharmaceutical Sciences, 1-1-20, Shin-machi, Nishi-Tokyo City, Tokyo 202-8585, Japan. Electronic address: motsuka@musashino-u.ac.jp.
Int J Pharm ; 565: 325-332, 2019 Jun 30.
Article em En | MEDLINE | ID: mdl-31075440
ABSTRACT
It is important to quantify amorphous solid in active pharmaceutical ingredients (API) of formulations to guarantee high-quality pharmaceutical products. In this study, we aimed to quantify amorphous solid in API by transmission Raman spectroscopy (TRS) and transmission near-infra-red spectroscopy (NIRS). To quantify the amount of trace amorphous solids contained in the API of a formulation, calibration curves of tablets containing 40% or 3% solifenacin succinate (SLFS) in the pharmaceutical formulations were prepared. The tablets varied in the ratio of amorphous solids in the API. TRS and NIRS were carried out with the calibration tablets and partial least square regression (PLSR) analysis was performed. Good results were obtained by both TRS and NIRS with the calibration model containing 40% SLFS. On the contrary, with the calibration model containing 3% SLFS, the PLSR analysis results using the TRS data were better than those using the NIRS data. The low content calibration model based on TRS showed good results with R2 of 0.999, RMSECV of 1.236, and LOD of 0.12. TRS can be used to specifically detect trace amorphous solids contained in the API of a formulation.