Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study.
Mangano, Francesco Guido; Hauschild, Uli; Veronesi, Giovanni; Imburgia, Mario; Mangano, Carlo; Admakin, Oleg.
Afiliação
  • Mangano FG; Department of Prevention and Communal Dentistry, Sechenov First Moscow State Medical University, Moscow, Russia. francescoguidomangano@gmail.com.
  • Hauschild U; Department of Post-graduate Education, Faculty of Oral and Dental Medicine, J.W. Goethe University, Frankfurt, Germany.
  • Veronesi G; Department of Medicine and Surgery, Research Center in Epidemiology and Preventive Medicine, University of Varese, Varese, Italy.
  • Imburgia M; Private Practice, Palermo, Italy.
  • Mangano C; Department of Dental Sciences, Vita and Salute University San Raffaele, Milan, Italy.
  • Admakin O; Department of Prevention and Communal Dentistry, Sechenov First Moscow State Medical University, Moscow, Russia.
BMC Oral Health ; 19(1): 101, 2019 06 06.
Article em En | MEDLINE | ID: mdl-31170969
ABSTRACT

BACKGROUND:

Until now, a few studies have addressed the accuracy of intraoral scanners (IOSs) in implantology. Hence, the aim of this in vitro study was to assess the accuracy of 5 different IOSs in the impressions of single and multiple implants, and to compare them.

METHODS:

Plaster models were prepared, representative of a partially edentulous maxilla (PEM) to be restored with a single crown (SC) and a partial prosthesis (PP), and a totally edentulous maxilla (TEM) to be restored with a full-arch (FA). These models were scanned with a desktop scanner, to capture reference models (RMs), and with 5 IOSs (CS 3600®, Trios3®, Omnicam®, DWIO®, Emerald®); 10 scans were taken for each model, using each IOS. All IOS datasets were loaded into a reverse-engineering software where they were superimposed on the corresponding RMs, to evaluate trueness, and superimposed on each other within groups, to determine precision. A statistical analysis was performed.

RESULTS:

In the SC, CS 3600® had the best trueness (15.2 ± 0.8 µm), followed by Trios3® (22.3 ± 0.5 µm), DWIO® (27.8 ± 3.2 µm), Omnicam® (28.4 ± 4.5 µm), Emerald® (43.1 ± 11.5 µm). In the PP, CS 3600® had the best trueness (23 ± 1.1 µm), followed by Trios3® (28.5 ± 0.5 µm), Omnicam® (38.1 ± 8.8 µm), Emerald® (49.3 ± 5.5 µm), DWIO® (49.8 ± 5 µm). In the FA, CS 3600® had the best trueness (44.9 ± 8.9 µm), followed by Trios3® (46.3 ± 4.9 µm), Emerald® (66.3 ± 5.6 µm), Omnicam® (70.4 ± 11.9 µm), DWIO® (92.1 ± 24.1 µm). Significant differences were found between the IOSs; a significant difference in trueness was found between the contexts (SC vs. PP vs. FA). In the SC, CS 3600® had the best precision (11.3 ± 1.1 µm), followed by Trios3® (15.2 ± 0.8 µm), DWIO® (27.1 ± 10.7 µm), Omnicam® (30.6 ± 3.3 µm), Emerald® (32.8 ± 10.7 µm). In the PP, CS 3600® had the best precision (17 ± 2.3 µm), followed by Trios3® (21 ± 1.9 µm), Emerald® (29.9 ± 8.9 µm), DWIO® (34.8 ± 10.8 µm), Omnicam® (43.2 ± 9.4 µm). In the FA, Trios3® had the best precision (35.6 ± 3.4 µm), followed by CS 3600® (35.7 ± 4.3 µm), Emerald® (61.5 ± 18.1 µm), Omnicam® (89.3 ± 14 µm), DWIO® (111 ± 24.8 µm). Significant differences were found between the IOSs; a significant difference in precision was found between the contexts (SC vs. PP vs. FA).

CONCLUSIONS:

The IOSs showed significant differences between them, both in trueness and in precision. The mathematical error increased in the transition from SC to PP up to FA, both in trueness than in precision.