Biomechanical In Vitro Test of a Novel Dynamic Spinal Stabilization System Incorporating Polycarbonate Urethane Material Under Physiological Conditions.
Beckmann, Agnes; Nicolini, Luis Fernando; Grevenstein, David; Backes, Hermann; Oikonomidis, Stavros; Sobottke, Rolf; Kobbe, Philipp; Hildebrand, Frank; Stoffel, Marcus; Markert, Bernd; Siewe, Jan; Herren, Christian.
Afiliação
  • Beckmann A; Institute of General Mechanics, RWTH Aachen University, Templergraben 64, Aachen 52062, Germany.
  • Nicolini LF; Institute of General Mechanics, RWTH Aachen University, Templergraben 64, Aachen 52062, Germany.
  • Grevenstein D; Centre for Orthopedic and Trauma Surgery, University of Cologne, Kerpener Street 62, Köln 50937, Germany.
  • Backes H; NGMedical GmbH, Morschborn 28, Nonnweiler-Primstal 66620, Germany.
  • Oikonomidis S; Centre for Orthopedic and Trauma Surgery, University of Cologne, Kerpener Street 62, Köln 50937, Germany.
  • Sobottke R; Rhein-Maas Klinik, Department of Orthopaedics and Trauma Surgery, Mauerfeldchen 25, Würselen 52146, Germany.
  • Kobbe P; Department for Trauma and Reconstructive Surgery, University Hospital RWTH Aachen, Pauwelsstraße 30, Aachen 52074, Germany.
  • Hildebrand F; Department for Trauma and Reconstructive Surgery, University Hospital RWTH Aachen, Pauwelsstraße 30, Aachen 52074, Germany.
  • Stoffel M; Institute of General Mechanics, RWTH Aachen University, Templergraben 64, Aachen 52062, Germany.
  • Markert B; Institute of General Mechanics, RWTH Aachen University, Templergraben 64, Aachen 52062, Germany.
  • Siewe J; Klinikum Leverkusen gGmbH, Clinic for Orthopedic and Trauma Surgery, Am Gesundheitspark 11, Leverkusen 51375, Germany.
  • Herren C; Department for Trauma and Reconstructive Surgery, University Hospital RWTH Aachen, Pauwelsstraße 30, Aachen 52074, Germany.
J Biomech Eng ; 142(1)2020 01 01.
Article em En | MEDLINE | ID: mdl-31314885
Posterior dynamic stabilization systems (PDSS) were developed to provide stabilization to pathologic or hypermobile spinal segments while maintaining the healthy biomechanics of the spine. Numerous novel dynamic devices incorporate the temperature and moisture dependent material polycarbonate urethane (PCU) due to its mechanical properties and biocompatibility. In this study, standardized pure moment in vitro tests were carried out on human lumbar spines to evaluate the performance of a device containing PCU. An environmental chamber with controlled moisture and temperature was included in the setup to meet the requirements of testing under physiological conditions. Three test conditions were compared: (1) native spine, (2) dynamic instrumentation, and (3) dynamic instrumentation with decompression. The ranges of motion, centers of rotation, and relative pedicle screw motions were evaluated. The device displayed significant stiffening in flexion-extension, lateral bending, and axial rotation load directions. A reduction of the native range of motion diminished the stiffening effect along the spinal column and has the potential to reduce the risk of the onset of degeneration of an adjacent segment. In combination with decompression, the implant decreased the native range of motion for flexion-extension and skew bending, but not for lateral bending and axial rotation. Curve fittings using the sigmoid function were performed to parameterize all load-deflection curves in order to enhance accurate numerical model calibrations and comparisons. The device caused a shift of the center of rotation (COR) in the posterior and caudal direction during flexion-extension loading.