Carbon microparticles as a physical carrier for ETV-ICP-MS.
Patocka, Jan; Cernohorský, Tomás; Krejcová, Anna; Slang, Stanislav.
Afiliação
  • Patocka J; Department of Environmental and Chemical Engineering, University of Pardubice, CZ-532 10, Pardubice, Czech Republic. Electronic address: st30721@student.upce.cz.
  • Cernohorský T; Department of Environmental and Chemical Engineering, University of Pardubice, CZ-532 10, Pardubice, Czech Republic.
  • Krejcová A; Department of Environmental and Chemical Engineering, University of Pardubice, CZ-532 10, Pardubice, Czech Republic.
  • Slang S; Center of Materials and Nanotechnologies - CEMNAT, University of Pardubice, CZ-532 10, Pardubice, Czech Republic.
Talanta ; 204: 555-560, 2019 Nov 01.
Article em En | MEDLINE | ID: mdl-31357333
ABSTRACT
In this study, carbon microparticles (CMs) as a physical carrier to enhance the analyte transport efficiency for inductively coupled plasma mass spectrometry connected with electrothermal vaporization (ETV-ICP-MS) are proposed. Carbon microparticles mixed with samples or calibration standard solutions were dosed as a slurry into the graphite furnace. The optimization of working conditions was done for the standard solution 0.5 µg L-1 of Au and Tl. The pyrolysis and vaporization temperatures were 500 and 2700 °C for Au and 400 and 1900 °C for Tl. The optimized CMs concentration was 1 g L-1 for Au and 2.5 g L-1 for Tl. For the quantification, external calibration standard solutions were used. The result obtained for Au in the digested CRM GBW 07601 (the informatory value 2.1 ±â€¯2 µg kg-1) was 2.2 ±â€¯0.1 µg kg-1. The result for Tl in the digested CRM BCR 679 (the certified value 3 ±â€¯0.3 µg kg-1) was 3.2 ±â€¯0.2 µg kg-1 and in the slurry 2.7 ±â€¯0.1 µg kg-1. The result for Tl in the CRM GBW 10052 (the certified value 57 ±â€¯11 µg kg-1) was 51 ±â€¯3 µg kg-1. The instrumental limits of detection were 0.016 ng L-1 for Au and 0.026 ng L-1 for Tl. The recoveries and repeatabilities measured on calibration standards were in the range 99-100% and 0.2-2.3% for Au and 100-111% and 2.9-6.7% for Tl.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2019 Tipo de documento: Article