Josephson Arbitrary Waveform Synthesizer as a Reference Standard for the Measurement of the Phase of Harmonics in Distorted Waveforms.
Georgakopoulos, Dimitrios; Budovsky, Ilya; Benz, Samuel P; Gubler, Gleb.
Afiliação
  • Georgakopoulos D; National Measurement Institute Australia, West Lindfield, NSW 2070, Australia.
  • Budovsky I; National Measurement Institute Australia, West Lindfield, NSW 2070, Australia.
  • Benz SP; National Institute of Standards and Technology, Boulder, CO 80305 USA.
  • Gubler G; D.I. Mendeleyev Institute for Metrology, 190005 St Petersburg, Russia.
Article em En | MEDLINE | ID: mdl-31579270
We have used the Josephson arbitrary waveform synthesizer (JAWS) to provide traceability for the phase of the harmonics, relative to their fundamental frequency, of a distorted waveform. For distorted waveforms with rms values from 0.154 to 0.2 V and harmonic magnitudes from 5% to 40% of the fundamental, our system can generate odd harmonics up to the 39th with best phase uncertainties from 0.001° to 0.010° (k = 2.0), depending on the harmonic number and harmonic magnitude. We anticipate that the ability of the JAWS to generate distorted waveforms with the lowest possible uncertainty in the magnitude, and phase spectra will make it a unique tool for low-frequency spectrum analysis.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2018 Tipo de documento: Article