Successful recovery from transcatheter aortic valve embolization by thoracic endovascular stent-graft repair.
Gen Thorac Cardiovasc Surg
; 68(8): 851-854, 2020 Aug.
Article
em En
| MEDLINE
| ID: mdl-31625083
We report a case of an 84-year-old Japanese female patient with transcatheter valve embolization treated with thoracic endovascular stenting. She was diagnosed with severe aortic stenosis and referred to our hospital. Transapical transcatheter aortic valve implantation (TA-TAVI) was selected because she had multiple comorbidities and peripheral vascular disease. During TA-TAVI, the transcatheter valve (TV) came loose and lodged in the distal aortic arch. The TV was floating and unstable, thereby affecting her hemodynamics. After a second TV was placed in the appropriate position with no trouble, we tried fixing the TV using a thoracic endovascular stent-graft. The procedure was successful, and the patient recovered well. The cause of embolization seemed to be associated with a relatively mild calcification of the native aortic valve. Although this complication is rare, several recovery procedures should be prepared.