Laser fibre, rather than the stone, may harm the scope: retrospective monocentric analysis of 26 pre- and intraoperative factors of flexible ureteroscope (fURS) damage.
Ozimek, Tomasz; Cordes, Jens; Gilbert, Nils; Hupe, Marie C; Wiessmeyer, Judith R; Schneider, Michael H; Merseburger, Axel S; Kramer, Mario Wolfgang.
Afiliação
  • Ozimek T; Department of Urology, University Hospital Schleswig-Holstein, Campus Luebeck, Ratzeburger Allee 160, 23538, Lübeck, Germany.
  • Cordes J; Department of Urology, University Hospital Schleswig-Holstein, Campus Luebeck, Ratzeburger Allee 160, 23538, Lübeck, Germany.
  • Gilbert N; Department of Urology, University Hospital Schleswig-Holstein, Campus Luebeck, Ratzeburger Allee 160, 23538, Lübeck, Germany.
  • Hupe MC; Department of Urology, University Hospital Schleswig-Holstein, Campus Luebeck, Ratzeburger Allee 160, 23538, Lübeck, Germany.
  • Wiessmeyer JR; Department of Urology, University Hospital Schleswig-Holstein, Campus Luebeck, Ratzeburger Allee 160, 23538, Lübeck, Germany.
  • Schneider MH; Department of Urology, University Hospital Schleswig-Holstein, Campus Luebeck, Ratzeburger Allee 160, 23538, Lübeck, Germany.
  • Merseburger AS; Department of Urology, University Hospital Schleswig-Holstein, Campus Luebeck, Ratzeburger Allee 160, 23538, Lübeck, Germany.
  • Kramer MW; Department of Urology, University Hospital Schleswig-Holstein, Campus Luebeck, Ratzeburger Allee 160, 23538, Lübeck, Germany. mario.kramer@uksh.de.
World J Urol ; 38(8): 2035-2040, 2020 Aug.
Article em En | MEDLINE | ID: mdl-31659464
ABSTRACT

PURPOSE:

The increasing number of flexible ureteroscopy procedures, the fragility of devices and their repair costs are a burden for urological departments worldwide. The objective was to investigate the impact of 26 pre- and intraoperative factors on reusable flexible ureteroscope (fURS) damage.

METHODS:

All procedures were conducted with reusable fURS Karl Storz Flex-X2 or Olympus URF-V. Statistical analysis was performed in RStudio (1.0.136) with Chi-square test and Mann-Whitney U tests (MWU).

RESULTS:

In total, 416 flexible ureteroscopies, performed between September 2013 and June 2017, were analysed. 283 (68.03%) of these were for kidney stone surgery, and 133 (31.97%) for diagnostic purposes. In total, 39 (9.38%) devices were postoperatively deemed defective. The application of reusable laser fibre through fURS was more common in cases with documented defects [17/39 (43.59%) vs. 102/377 (27.06%), p = 0.047]. Other factors such as application of nitinol basket, biopsy via fURS, insertion of access sheath (UAS), as well as stone burden [median kidney stone maximal diameter 6 mm (min 2.0; max 30.0) vs. 6 mm (min 1.0 vs. max 30.0)] showed no influence on fURS damage rate (p > 0.05). The infundibulopelvic angle (IPA) was steeper in cases with fURS damage as compared to cases without damage [median 44.0° (min 20.0; max 81.0) vs. 55.0 (min 7.0; max 122.0), p < 0.001].

CONCLUSIONS:

Application of laser fibre via fURS can be considered as a risk factor of fURS damage. Stone burden, as well as the usage of not-sharp ended devices as nitinol baskets or forceps, is primarily not responsible for fURS damage.