Simultaneous determination of di(2-ethylhexyl) phthalate and diisononylcyclohexane-1,2-dicarboxylate and their monoester metabolites in four labile blood products by liquid chromatography tandem mass spectrometry.
J Pharm Biomed Anal
; 181: 113063, 2020 Mar 20.
Article
em En
| MEDLINE
| ID: mdl-31927338
Di(2-ethylhexyl) phthalate (DEHP) is a common plasticizer that is largely used for PVC blood bags. The migration of DEHP from medical devices into labile blood products (LBP) is a well-known situation. While DEHP has beneficial effects on the storage of red blood cells, it can have toxicological impact due to its potential reprotoxic effects (classified group 1B). Since July 1st, 2015, the French law prohibits the use of tubing made in DEHP-plasticized PVC in paediatric, neonatal and maternity wards. This provision, which could extend in several years more widely to medical devices used for drugs infusion, dialysis, feeding and blood bags, has led manufacturers to replace DEHP to alternative plasticizers such as diisononylcyclohexane-1,2-dicarboxylate (DINCH). In this paper, a liquid chromatography-tandem mass spectrometry (LCMS/MS) method has been developed and validated for the determination of DEHP, DINCH and their corresponding monoester metabolites (MEHP and MINCH) in four labile blood products (LBP): whole blood (WB), red cells concentrate (RCC), plasma and platelet concentrate (PC). Due to strong contamination of blank LBP by DEHP because of its ubiquitous presence in working environment and despite the attention paid to avoid contamination of solvents and glassware, a trap chromatographic column was implemented between the solvent mixing chamber and the injector of the LC system. This set-up permitted to discriminate DEHP present in the sample to DEHP brought by the environmental contamination. In the optimized conditions, all compounds were separated in less than 10â¯min. The analytes were extracted from LBP samples using a liquid-liquid extraction. After optimization, recoveries were ranged from 47 to 96 %, depending on the analytes and the nature of LBP. Except for DEHP which exhibited RSD values of intermediate precision higher than 20 % at a concentration of 25â¯nM, all the precision results (repeatability and intermediate precision) were lower than 16 % and trueness values ranged from -16.2-19.8%. Using the validated method, the leachability of DEHP and DINCH from corresponding PVC-blood bags was investigated and the concentrations of their corresponding metabolites, MEHP and MINCH, were determined in whole blood, red cells concentrate, plasma and platelet concentrate.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Plastificantes
/
Preservação de Sangue
/
Ácidos Cicloexanocarboxílicos
/
Ácidos Dicarboxílicos
/
Dietilexilftalato
/
Extração Líquido-Líquido
Idioma:
En
Ano de publicação:
2020
Tipo de documento:
Article