T1 Relaxation Time of Achilles Tendon at 3 Tesla with Special Reference to Relevant Clinical Score: A Preliminary Study.
Tbini, Zeineb; Mars, Mokhtar; Bouaziz, Mouna.
Afiliação
  • Tbini Z; Laboratory of Research in Biophysics and Medical Technologies, Higher Institute of Medical Technologies of Tunis, El Manar University, 09, Rue Doctor Zouheir Safi -1006 Tunis, Tunisia.
  • Mars M; Laboratory of Research in Biophysics and Medical Technologies, Higher Institute of Medical Technologies of Tunis, El Manar University, 09, Rue Doctor Zouheir Safi -1006 Tunis, Tunisia.
  • Bouaziz M; Radiology Department, Mohamed Kassab Institute of Orthopedics, 2010 Ksar Said, Tunisia.
Curr Med Imaging Rev ; 16(2): 164-173, 2020.
Article em En | MEDLINE | ID: mdl-32003317
ABSTRACT

PURPOSE:

The purpose of this study was to investigate T1 relaxation time of the human Achilles tendon, to test its short-term repeatability as well as the minimal detectable change, and to assess the extent that correlate with clinical symptoms.

METHODS:

Twenty asymptomatic volunteers and eighteen patients with clinically and sonographically confirmed tendinopathy were scanned for ankle using a 3 Tesla (T) MR scanner. T1 maps were calculated from a variable flip angle gradient echo Ultra-short echo time sequence (VFA-GE UTE) and inversion recovery spin echo sequence (IR-SE) using a self-developed matlab algorithm in three regions of interest of Achilles Tendon (AT). Signal to Noise Ratio (SNR) between the two sequences was evaluated. INTRA-class Correlation Coefficient (ICC), Coefficient of Variation (CV) and the Least Significant Change (LSC) were calculated, to test short-term repeatability of T1. Subjects were assessed by the VISA-A clinical score. P values less than 0.005 were considered statistically significant.

RESULTS:

Mean T1 values were 427.09 ± 53.37 ms and 528.70 ± 103.50 ms using IR-SE sequence and 575.43 ± 110.60 ms and 875.81 ± 425.77 ms with VFA-GE UTE sequence in the whole AT for volunteers and patients, respectively. T1 values showed a significant difference between volunteers and patients (P=0.001). Regional variation of T1 in healthy and tendinopathic AT were greater for VFA-GE UTE sequence than for IR-SE sequence. VFA-GE UTE sequence showed clearly higher SNR compared to IR-SE sequence. Short-term repeatability of T1 values for volunteers showed an LSC of 22% and 14% for IR-SE sequence and VFA-GE UTE sequence, respectively. For patients, LSC was 14% and 5% for IR-SE sequence and VFA-GE UTE sequence, respectively. There was no correlation between T1 and VISA-A clinical score (p>0.005).

CONCLUSION:

VFA-GE UTE sequence used for T1 mapping calculation demonstrated short acquisition time and clearly high SNR. Results revealed that T1 relaxation time can be used as a biomarker to differentiate between healthy and pathologic Achilles tendon. However, T1 showed no correlation with the VISA-A clinical score.