CT-based radiomics for preoperative prediction of early recurrent hepatocellular carcinoma: technical reproducibility of acquisition and scanners.
Hu, Hang-Tong; Shan, Quan-Yuan; Chen, Shu-Ling; Li, Bin; Feng, Shi-Ting; Xu, Er-Jiao; Li, Xin; Long, Jian-Yan; Xie, Xiao-Yan; Lu, Ming-de; Kuang, Ming; Shen, Jing-Xian; Wang, Wei.
Afiliação
  • Hu HT; Department of Medical Ultrasonics, Institute of Diagnostic and Interventional Ultrasound, Ultrasomics Artificial Intelligence X-Lab, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhong Shan Road 2, Guangzhou, 510080, China.
  • Shan QY; Department of Medical Ultrasonics, Institute of Diagnostic and Interventional Ultrasound, Ultrasomics Artificial Intelligence X-Lab, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhong Shan Road 2, Guangzhou, 510080, China.
  • Chen SL; Department of Medical Ultrasonics, Institute of Diagnostic and Interventional Ultrasound, Ultrasomics Artificial Intelligence X-Lab, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhong Shan Road 2, Guangzhou, 510080, China.
  • Li B; Clinical trials Unit, The First Affiliated Hospital of Sun Yat-Sen University, Guangzhou, China.
  • Feng ST; Department of Radiology, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhong Shan Road 2, Guangzhou, 510080, China.
  • Xu EJ; Department of Medical Ultrasonics, The Third Affiliated Hospital of Sun Yat-Sen University, Guangdong Key Laboratory of Liver Disease Research, Guangzhou, 510630, Guangdong Province, China.
  • Li X; GE Healthcare, Shanghai, 200030, China.
  • Long JY; Clinical trials Unit, The First Affiliated Hospital of Sun Yat-Sen University, Guangzhou, China.
  • Xie XY; Department of Medical Ultrasonics, Institute of Diagnostic and Interventional Ultrasound, Ultrasomics Artificial Intelligence X-Lab, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhong Shan Road 2, Guangzhou, 510080, China.
  • Lu MD; Department of Medical Ultrasonics, Institute of Diagnostic and Interventional Ultrasound, Ultrasomics Artificial Intelligence X-Lab, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhong Shan Road 2, Guangzhou, 510080, China.
  • Kuang M; Department of Liver Surgery, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhong Shan Road 2, Guangzhou, 510080, China.
  • Shen JX; Department of Medical Ultrasonics, Institute of Diagnostic and Interventional Ultrasound, Ultrasomics Artificial Intelligence X-Lab, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhong Shan Road 2, Guangzhou, 510080, China.
  • Wang W; Department of Liver Surgery, The First Affiliated Hospital of Sun Yat-Sen University, 58 Zhong Shan Road 2, Guangzhou, 510080, China.
Radiol Med ; 125(8): 697-705, 2020 Aug.
Article em En | MEDLINE | ID: mdl-32200455
ABSTRACT

PURPOSE:

To test the technical reproducibility of acquisition and scanners of CT image-based radiomics model for early recurrent hepatocellular carcinoma (HCC).

METHODS:

We included primary HCC patient undergone curative therapies, using early recurrence as endpoint. Four datasets were constructed 109 images from hospital #1 for training (set 1 1-mm image slice thickness), 47 images from hospital #1 for internal validation (sets 2 and 3 1-mm and 10-mm image slice thicknesses, respectively), and 47 images from hospital #2 for external validation (set 4 vastly different from training dataset). A radiomics model was constructed. Radiomics technical reproducibility was measured by overfitting and calibration deviation in external validation dataset. The influence of slice thickness on reproducibility was evaluated in two internal validation datasets.

RESULTS:

Compared with set 1, the model in set 2 indicated favorable prediction efficiency (the area under the curve 0.79 vs. 0.80, P = 0.47) and good calibration (unreliability statistic U P = 0.33). However, in set 4, significant overfitting (0.63 vs. 0.80, P < 0.01) and calibration deviation (U P < 0.01) were observed. Similar poor performance was also observed in set 3 (0.56 vs. 0.80, P = 0.02; U P < 0.01).

CONCLUSIONS:

CT-based radiomics has poor reproducibility between centers. Image heterogeneity, such as slice thickness, can be a significant influencing factor.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X / Carcinoma Hepatocelular / Neoplasias Hepáticas / Recidiva Local de Neoplasia Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Interpretação de Imagem Radiográfica Assistida por Computador / Tomografia Computadorizada por Raios X / Carcinoma Hepatocelular / Neoplasias Hepáticas / Recidiva Local de Neoplasia Idioma: En Ano de publicação: 2020 Tipo de documento: Article