Your browser doesn't support javascript.
loading
A case of cardiac strangulation following epicardial pacemaker implantation.
Miyagi, Chihiro; Ochiai, Yoshie; Ando, Yusuke; Hisahara, Manabu; Baba, Hironori; Takigawa, Tomoya; Tokunaga, Shigehiko.
Afiliação
  • Miyagi C; Department of Cardiovascular Surgery, Japan Community Health Care Organization (JCHO), Kyushu Hospital, 1-8-1, Kishinoura, Yahatanishi-ku, Kitakyushu, 806-8501, Japan.
  • Ochiai Y; Department of Cardiovascular Surgery, Japan Community Health Care Organization (JCHO), Kyushu Hospital, 1-8-1, Kishinoura, Yahatanishi-ku, Kitakyushu, 806-8501, Japan. yoshie558@yahoo.co.jp.
  • Ando Y; Department of Cardiovascular Surgery, Japan Community Health Care Organization (JCHO), Kyushu Hospital, 1-8-1, Kishinoura, Yahatanishi-ku, Kitakyushu, 806-8501, Japan.
  • Hisahara M; Department of Cardiovascular Surgery, Japan Community Health Care Organization (JCHO), Kyushu Hospital, 1-8-1, Kishinoura, Yahatanishi-ku, Kitakyushu, 806-8501, Japan.
  • Baba H; Department of Cardiovascular Surgery, Japan Community Health Care Organization (JCHO), Kyushu Hospital, 1-8-1, Kishinoura, Yahatanishi-ku, Kitakyushu, 806-8501, Japan.
  • Takigawa T; Department of Cardiovascular Surgery, Japan Community Health Care Organization (JCHO), Kyushu Hospital, 1-8-1, Kishinoura, Yahatanishi-ku, Kitakyushu, 806-8501, Japan.
  • Tokunaga S; Department of Cardiovascular Surgery, Japan Community Health Care Organization (JCHO), Kyushu Hospital, 1-8-1, Kishinoura, Yahatanishi-ku, Kitakyushu, 806-8501, Japan.
Gen Thorac Cardiovasc Surg ; 68(12): 1499-1502, 2020 Dec.
Article em En | MEDLINE | ID: mdl-32270437
ABSTRACT
An 8-year-old boy had undergone permanent epicardial pacemaker implantation with a Y-shaped bipolar ventricular lead on day 6 after birth for treatment of congenital complete atrioventricular block. He was found to have pulmonary stenosis and mitral stenosis by follow-up echocardiography. Further studies including computed tomography and cardiac catheterization revealed that the pacemaker lead had completely encircled the cardiac silhouette and was in a state of "cardiac strangulation". We removed the previous pacing leads and generator and implanted a new epicardial dual-chamber pacing system in the right atrium and right ventricle. Additionally, an expanded polytetrafluoroethylene sheet was placed between the new leads and the heart to prevent recurrence of cardiac strangulation.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial Idioma: En Ano de publicação: 2020 Tipo de documento: Article