Your browser doesn't support javascript.
loading
Inverse Laplace transform and multiexponential fitting analysis of T2 relaxometry data: a phantom study with aqueous and fat containing samples.
Ioannidis, Georgios S; Nikiforaki, Katerina; Kalaitzakis, Georgios; Karantanas, Apostolos; Marias, Kostas; Maris, Thomas G.
Afiliação
  • Ioannidis GS; Foundation for Research and Technology-Hellas (FORTH), Institute of Computer Science (ICS), Computational Bio-Medicine Laboratory (CBML), N.Plastira 100, Vassilika Vouton, Heraklion, GR-70013, Crete, Greece. grs.ioannidis@gmail.com.
  • Nikiforaki K; School of Medicine, University of Crete, Heraklion, Greece. grs.ioannidis@gmail.com.
  • Kalaitzakis G; Foundation for Research and Technology-Hellas (FORTH), Institute of Computer Science (ICS), Computational Bio-Medicine Laboratory (CBML), N.Plastira 100, Vassilika Vouton, Heraklion, GR-70013, Crete, Greece.
  • Karantanas A; School of Medicine, University of Crete, Heraklion, Greece.
  • Marias K; School of Medicine, University of Crete, Heraklion, Greece.
  • Maris TG; Department of Medical Physics, University of Crete, Heraklion, Greece.
Eur Radiol Exp ; 4(1): 28, 2020 05 07.
Article em En | MEDLINE | ID: mdl-32378090
ABSTRACT

BACKGROUND:

The inverse Laplace transform (ILT) is the most widely used method for T2 relaxometry data analysis. This study examines the qualitative agreement of ILT and a proposed multiexponential (Mexp method) regarding the number of T2 components. We performed a feasibility study for the voxelwise characterisation of heterogeneous tissue with T2 relaxometry.

METHODS:

Eleven samples of aqueous, fatty and mixed composition were analysed using ILT and Mexp. The phantom was imaged using a 1.5-T system with a single slice T2 relaxometry 25-echo Carr-Purcell-Meiboom-Gill sequence in order to obtain the T2 decay curve with 25 equidistant echo times. The adjusted R2 goodness of fit criterion was used to determine the number of T2 components using the Mexp method on a voxel-based analysis. Comparison of mean and standard deviation of T2 values for both methods was performed by fitting a Gaussian function to the ILT resulting vector.

RESULTS:

Phantom results showed pure monoexponential decay for acetone and water and pure biexponential behaviour for corn oil, egg yolk, and 35% fat milk cream, while mixtures of egg whites and yolks as well as milk creams with 12-20% fatty composition exhibit mixed monoexponential and biexponential behaviour at different fractions. The number of T2 components by the Mexp method was compared to the ILT-derived spectrum as ground truth.

CONCLUSIONS:

Mexp analysis with the adjusted R2 criterion can be used for the detection of the T2 distribution of aqueous, fatty and mixed samples with the added advantage of voxelwise mapping.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética Idioma: En Ano de publicação: 2020 Tipo de documento: Article