Image quality in dual-source multiphasic dynamic computed tomography of the abdomen: evaluating the effects of a low tube voltage (70 kVp) in combination with contrast dose reduction.
Miyoshi, Keisuke; Onoda, Hideko; Tanabe, Mashiro; Nakao, Sei; Higashi, Mayumi; Iida, Etsushi; Okada, Munemasa; Furukawa, Matakazu; Ito, Katsuyoshi.
Afiliação
  • Miyoshi K; Department of Radiology, Yamaguchi University Graduate School of Medicine, 1-1-1 Minami-Kogushi, Ube, Yamaguchi, 755-8505, Japan.
  • Onoda H; Department of Radiology, Yamaguchi University Graduate School of Medicine, 1-1-1 Minami-Kogushi, Ube, Yamaguchi, 755-8505, Japan.
  • Tanabe M; Department of Radiology, Yamaguchi University Graduate School of Medicine, 1-1-1 Minami-Kogushi, Ube, Yamaguchi, 755-8505, Japan.
  • Nakao S; Department of Radiology, Yamaguchi University Graduate School of Medicine, 1-1-1 Minami-Kogushi, Ube, Yamaguchi, 755-8505, Japan.
  • Higashi M; Department of Radiology, Yamaguchi University Graduate School of Medicine, 1-1-1 Minami-Kogushi, Ube, Yamaguchi, 755-8505, Japan.
  • Iida E; Department of Radiology, Yamaguchi University Graduate School of Medicine, 1-1-1 Minami-Kogushi, Ube, Yamaguchi, 755-8505, Japan.
  • Okada M; Department of Radiology, Yamaguchi University Graduate School of Medicine, 1-1-1 Minami-Kogushi, Ube, Yamaguchi, 755-8505, Japan.
  • Furukawa M; Department of Radiology, Yamaguchi University Graduate School of Medicine, 1-1-1 Minami-Kogushi, Ube, Yamaguchi, 755-8505, Japan.
  • Ito K; Department of Radiology, Yamaguchi University Graduate School of Medicine, 1-1-1 Minami-Kogushi, Ube, Yamaguchi, 755-8505, Japan. itokatsu@yamaguchi-u.ac.jp.
Abdom Radiol (NY) ; 45(11): 3755-3762, 2020 11.
Article em En | MEDLINE | ID: mdl-32382819
ABSTRACT

PURPOSE:

To compare the image quality of multiphasic (arterial, portal, and equilibrium phases) dynamic computed tomography (CT) of the abdomen obtained by a low tube voltage (70kVp) in combination with a half-dose iodine load using low-concentration contrast agent in high tube output dual-source CT with a standard tube voltage (120kVp) and full-dose iodine load using the same group of adult patients.

METHODS:

Fifty-five patients who underwent both low-tube-voltage (70kVp) abdominal CT with a half-dose iodine load and standard-tube-voltage (120kVp) CT with a full-dose iodine load were analyzed. The mean CT values and signal-to-noise ratio (SNR) of the liver, aorta and portal veins were quantitatively assessed. In addition, the contrast enhancement of the abdominal organs and overall image quality were qualitatively evaluated.

RESULTS:

The mean CT values and SNR of the liver parenchyma were significantly higher in 70-kVp protocol than in 120-kVp protocol in all 3 phases (p = 0.018 ~ < 0.001). Regarding the qualitative analysis, the overall image quality in the 70-kVp protocol was significantly better than in the 120-kVp protocol in all 3 phases (p < 0.001). In addition, the contrast enhancement scores of the liver parenchyma and hepatic vein in the equilibrium phase were also significantly higher in the 70-kVp protocol than in the 120-kVp protocol (p < 0.001).

CONCLUSION:

A low tube voltage (70kVp) in combination with a half-dose iodine load using a low-concentration contrast agent and an iterative reconstruction algorithm in high tube output dual-source CT may improve the contrast enhancement and image quality in multiphasic dynamic CT of the abdomen in patients under 71 kg of body weight.