Your browser doesn't support javascript.
loading
[Preliminary study of a mask filtration test device with respiratory simulation].
Nie, X Q; Fan, L Y; Chen, Q; Qin, J Y; Luo, Y Y; Chen, W H.
Afiliação
  • Nie XQ; Department of Occupational and Environmental Health, School of Public Health, Tongji Medical College, Huazhong University of Science & Technology, Key Laboratory of the Ministry of Environment and Health, Wuhan 430030, China.
  • Fan LY; Department of Occupational and Environmental Health, School of Public Health, Tongji Medical College, Huazhong University of Science & Technology, Key Laboratory of the Ministry of Environment and Health, Wuhan 430030, China.
  • Chen Q; Department of Occupational and Environmental Health, School of Public Health, Tongji Medical College, Huazhong University of Science & Technology, Key Laboratory of the Ministry of Environment and Health, Wuhan 430030, China.
  • Qin JY; Department of Occupational and Environmental Health, School of Public Health, Tongji Medical College, Huazhong University of Science & Technology, Key Laboratory of the Ministry of Environment and Health, Wuhan 430030, China.
  • Luo YY; Department of Occupational and Environmental Health, School of Public Health, Tongji Medical College, Huazhong University of Science & Technology, Key Laboratory of the Ministry of Environment and Health, Wuhan 430030, China.
  • Chen WH; Department of Occupational and Environmental Health, School of Public Health, Tongji Medical College, Huazhong University of Science & Technology, Key Laboratory of the Ministry of Environment and Health, Wuhan 430030, China.
Article em Zh | MEDLINE | ID: mdl-32447886
ABSTRACT

Objective:

To develop an air particulate protective mask filter test device that can simulate the dynamics process of human breathing.

Methods:

The new device used two air pumps working alternately to simulate the dynamics process of human breathing. On March 4th to 17th, 2017, the new device and the traditional one-way airflow mask filtration test device were used to measure the internal and external particle levels of 39 masks of 13 models of 6 brands, and then the filtration efficiency of the mask was calculated and the test results were compared.

Results:

For the mask without breathing valve, there was no statistically significant difference between the filter efficiency test results of the new device and the traditional unidirectional airflow filter performance test device (P>0.05) . For masks with breathing valves, the new device detected that three of them had lower filtration efficiency (99.50% vs 98.63%, P<0.01) . After sealing the mask breathing valve with glue, the filtering efficiency of the mask with a breathing valve detected by the new device significantly improved (98.63% vs 99.50%, P<0.01) .

Conclusion:

This new device can simulate the dynamic process of human exhalation and inhalation, and measure the filtration efficiency of the mask. For masks with breathing valves, the new device makes it easier to detect the decrease in the filtering efficiency of the mask caused by the breathing valve.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dispositivos de Proteção Respiratória / Desenho de Equipamento / Filtração / Máscaras Idioma: Zh Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dispositivos de Proteção Respiratória / Desenho de Equipamento / Filtração / Máscaras Idioma: Zh Ano de publicação: 2020 Tipo de documento: Article