Positioning error analysis of the fraxion localization system in the intracranial stereotactic radiotherapy of tumors.
He, D-C; Zhu, Z-J; Zhang, X-Y; Zhang, Y; Hong, J; Shi, T-T; Yu, C-H; Han, J-H.
Afiliação
  • He DC; Department of Radiation Oncology, The Affiliated Huaian No.1 People's Hospital of Nanjing Medical University, No. 1 Huanghe Road West, Huaiyin District, Huai'an, Jiangsu, 223300, People's Republic of China.
  • Zhu ZJ; Department of Radiation Oncology, The Affiliated Huaian No.1 People's Hospital of Nanjing Medical University, No. 1 Huanghe Road West, Huaiyin District, Huai'an, Jiangsu, 223300, People's Republic of China.
  • Zhang XY; Department of Radiation Oncology, The Affiliated Huaian No.1 People's Hospital of Nanjing Medical University, No. 1 Huanghe Road West, Huaiyin District, Huai'an, Jiangsu, 223300, People's Republic of China.
  • Zhang Y; Department of Radiation Oncology, The Affiliated Huaian No.1 People's Hospital of Nanjing Medical University, No. 1 Huanghe Road West, Huaiyin District, Huai'an, Jiangsu, 223300, People's Republic of China.
  • Hong J; Department of Radiation Oncology, The Affiliated Huaian No.1 People's Hospital of Nanjing Medical University, No. 1 Huanghe Road West, Huaiyin District, Huai'an, Jiangsu, 223300, People's Republic of China.
  • Shi TT; Department of Radiation Oncology, The Affiliated Huaian No.1 People's Hospital of Nanjing Medical University, No. 1 Huanghe Road West, Huaiyin District, Huai'an, Jiangsu, 223300, People's Republic of China.
  • Yu CH; Department of Radiation Oncology, The Affiliated Huaian No.1 People's Hospital of Nanjing Medical University, No. 1 Huanghe Road West, Huaiyin District, Huai'an, Jiangsu, 223300, People's Republic of China.
  • Han JH; Department of Radiation Oncology, The Affiliated Huaian No.1 People's Hospital of Nanjing Medical University, No. 1 Huanghe Road West, Huaiyin District, Huai'an, Jiangsu, 223300, People's Republic of China. hanjh220@163.com.
Clin Transl Oncol ; 23(1): 43-47, 2021 Jan.
Article em En | MEDLINE | ID: mdl-32462396
ABSTRACT

OBJECTIVE:

To investigate positioning error analysis of the Fraxion localization system in the intracranial stereotactic radiotherapy of tumors.

METHODS:

64 patients were divided into two groups a control group (36 patients with the standard thermoplastic mask) and a model group (28 patients with the Fraxion localization system). 3D images of the treated position were obtained by cone-beam computed tomography (CBCT). Positioning errors were obtained by, respectively, registering these two sets of CBCT images to planning CT images, using a 6°-freedom robotic patient positioning system (HexaPOD Evo RT System). The changes in positioning errors with the Fraxion localization system and with the standard thermoplastic mask were analyzed.

RESULTS:

CBCT scan results of the model group showed that the mean of linear error of three directions [superior-inferior (SI), lateral (LAT), and anterior-posterior (AP)] was 0.710 ± 0.676 mm, 0.817 ± 0.687 mm, and 0.710 ± 0.685 mm, respectively. The corresponding PTV was 1.23 mm, 1.26 mm, and 1.36 mm. The differences between the 3D images and the planned CT images were significant (p < 0.001).

CONCLUSION:

The Fraxion radiotherapy system can not only improve the positioning accuracy and reduce positioning errors but also narrow the PTV margin and reduce the radiated volume of normal tissue.